Chapter 31

Fundamentals of Measurement

A workpiece is considered good only if someone has measured it and can say, with good reason, that its dimensions fall within the tolerances. This chapter opens the section on metrology and lays its foundations: what it means to measure, where confidence in an instrument comes from, what errors are made and how they are reduced, how uncertainty is calculated and reported, when a single measurement is sufficient to accept or reject a workpiece, and why temperature, cleanliness, and the instrument’s position matter just as much as the instrument itself.

31.1 What does it mean to measure?

To measure means compare a quantity with its unit and determine how many times it fits (Chapter 2). In the workshop, it seems like a simple task: you place the caliper against the part and read a number. In reality, there are seven factors behind that number, and each one can introduce an error:

  • the quantity being measured: the precise value we wish to determine. “The diameter of the pin” is not enough: in which cross-section? between which points? at what temperature? A slightly oval pin has many different diameters;
  • the value: the number you get, which is never the true value, because no one knows the true value; you can only approximate it;
  • l'unit: in mechanics, the millimeter and the micrometer (µm), which are always specified;
  • the method: the method of measurement (direct, indirect, by comparison, paragraph 31.2) and the written procedure that describes how to perform it;
  • the instrument: caliper, micrometer, dial indicator, measuring machine, along with its specifications and calibration status;
  • the operator: the person who, through their skill, measures the force applied and interprets the results;
  • l'environment: temperature, cleanliness, vibrations, light.
Methoddirect, indirect, comparisonInstrument25–50 mm micrometerOperatorskill, strength, readingEnvironmenttemperature, cleanliness, lightingUnitmillimeter (SI)Measuringthe diameter of the pinResult: 25.012 ± 0.004 mm
Figure 31.1 · The result of a measurement depends on the quantity being measured, the method, the instrument, the operator, the environment, and the unit. If even one of these is neglected, the final number is meaningless.

A complete measurement result consists of three parts: value, uncertainty, and unit, for example, 25.012 ± 0.004 mm. The uncertainty (paragraph 31.7) indicates how reliable the value is. A measurement written without an uncertainty is only half the story: “25.012 mm” as read from a vernier caliper and from a coordinate measuring machine does not mean the same thing.

📐
Metrology. It is the science of measurement. The aspect relevant to the workshop is dimensional metrology: lengths, diameters, angles, shapes, threads, and roughness. The following chapters cover this topic instrument by instrument: linear measurements (Chapter 32), angles and cones (33), geometric tolerances (34), threading (Chapter 35), roughness (Chapter 36), modern measurement methods and measuring machines (Chapter 37), and quality control (Chapter 38).

31.2 Direct, Indirect, and Comparative Measurement

The same diameter can be measured in different ways. The method is chosen based on the required accuracy, the number of workpieces, and the available tools.

LiveThe tool providesdirectly the valueIndirectDCMeasure C and calculateD = C / πFor comparison50,00012I set the sample to zero,I read the difference
Figure 31.2 · On the left, the direct measurement; in the center, the indirect measurement (the diameter calculated from the circumference); on the right, the measurement by comparison: the comparator is zeroed on the block (1) and then reads the deviation of the workpiece (2).

Direct measurement

The instrument immediately provides the value of the measurement: the vernier caliper, the micrometer, and the ruler. This is the most common method. The instrument’s measuring range must cover the entire dimension; therefore, for a diameter of 80 mm, a micrometer with a range of 75–100 mm is required.

Indirect measurement

Other quantities are measured, and the desired value is calculated using a formula: the diameter of a shuttlecock from its circumference, the mean diameter of a thread using the three-thread method (Chapter 35), an angle using the sine bar (Chapter 33), and a radius from the chord and the arc. Errors in the initial measurements are carried over into the result, magnified or reduced depending on the formula.

Example 31.1 · The diameter of a shuttlecock measured with a tape measure

A flywheel is too large for any of the workshop's measuring instruments. Using a steel tape wrapped around the rim, the circumference is measured to be C = 1885.0 mm, with a measurement uncertainty of ±0.5 mm. What is the diameter, and what is the uncertainty?

D = C / π = 1885.0 / 3.1416 = 600.01 mm
The circumference error is also divided by π: 0.5 / 3.1416 = 0.16 mm
D = 600.01 ± 0.16 mm: The indirect method reduced the reading error by a factor of π

Comparison Measure (Differential)

The instrument does not measure the entire dimension, but only the difference between the workpiece and a known reference sample: a block, a stack of blocks, a ring, or a reference workpiece. The comparator is zeroed against the reference sample, and then the deviation of the workpiece is read. The comparator operates within a range of a few tenths of a millimeter, where it is most accurate, and many errors (temperature, deformation, stand errors) occur equally on both the reference sample and the workpiece and therefore cancel each other out. This is the method used in mass production and metrology laboratories.

Example 31.2 · A dimension derived by difference

The mounted thousandth-of-a-millimeter comparator is zeroed on a stack of blocks measuring 40 + 10 = 50.000 mm. When moved to the workpiece, the pointer reads −7 divisions of 0.001 mm.

dimension = sample + reading = 50.000 + (−0.007) = 49.993 mm
49.993 mm: the dial indicator measured only 7 µm; the block “knows” the rest

A special case of this comparison involves pass/fail gauges (Chapter 14): they do not provide a numerical value, but immediately indicate whether the workpiece is acceptable or must be rejected.

31.3 Traceability and Calibration Chain

Why should we believe what a micrometer tells us? Because someone in a laboratory compared it to a more precise standard; that standard was compared to an even better one, and so on, all the way up to the national standard. This property is called traceability (in English traceability): The result of a measurement is related to the International System of Units through a unbroken chain of calibrations, each with its own stated uncertainty.

SI Definition of the Meterspeed of lightINRiM, national standardStabilized lasers, TurinLAT Calibration Centeraccredited by ACCREDIACompany standardsblocks, rings, swabsWorkshop toolsmicrometers, calipers, dial indicatorsThe workpiecethe dimensions on the drawingthe uncertaintyincreases with eachstepEach ring is calibrated withthe one above it and has its owncertificate: the chainmust not be broken.
Figure 31.3 · The chain of traceability for length. As you move down the chain, the instruments become more numerous and more robust, but each step adds a little uncertainty.
  • At the top is the definition of the meter: the distance that light travels in a vacuum in 1/299,792,458 of a second. It is not an object, but a definition that anyone can verify with the right instruments.
  • In Italy, this is carried out by INRiM (the National Institute of Metrological Research, in Turin) using stabilized lasers and interferometers, which it uses to calibrate the laboratories’ reference sample blocks.
  • Accredited calibration centers (accredited in Italy by ACCREDIA and known as LAT centers) calibrate companies' samples and instruments, and issue certificates that are recognized internationally.
  • The company stores its reference standards (blocks, rings, swabs, and zeroing rods) and uses them to calibrate its workshop instruments.
  • Finally, the machine tool measures the workpiece.

If one link in the chain is missing (a micrometer that has never been calibrated, a gauge with an expired certificate, a sample “we trust”), the measurement is no longer traceable: it may be correct, but no one can prove it to a customer or an inspector. The standards on quality systems (Chapter 38) require precisely this: identified, calibrated, and traceable instruments.

31.4 The Characteristics of an Instrument

In an instrument's catalog and certificate, there are a few terms you should be familiar with, because they don't mean the same thing.

FeatureWhat does that mean?Example
Measurement rangethe range over which the instrument measuresmicrometer 25–50 mm; dial indicator 0–10 mm
Division Valuethe difference between two adjacent marks on the scale0.01 mm on the micrometer drum
Resolutionthe smallest change that the instrument can display; on digital instruments, it is the last digit0.01 mm digital caliper; 0.001 mm digital micrometer
Sensitivityhow much the reading changes for a unit change in the quantitycomparator: 157 mm dial for 1 mm of travel
Linearityto what extent the response remains proportional across the entire rangeerror that increases toward full scale
RepeatabilityHow similar are repeated measurements of the same quantity, under the same conditions?Ten readings that fit within 2 µm
Hysteresisthe difference between the reading obtained when approaching from below and the reading obtained when approaching from abovecomparator with friction and clearances in the gears
Maximum permissible errorthe error limit guaranteed by the manufacturer and the calibration check±3 µm micrometer across the entire field
⚠️
Resolution does not mean accuracy. A digital caliper displays hundredths, but its maximum permissible error is usually two or three hundredths. The third digit on the display is perfectly legible even when it is incorrect. The instrument is evaluated based on its certificate, not on the number of digits.
Example 31.3 · The Sensitivity of a Comparator

A centesimal comparator makes one full revolution of the pointer for every millimeter of travel; the dial has 100 divisions, and the tip of the pointer traces a circle with a radius of 25 mm.

One revolution of the tip = 2 · π · 25 = 157.08 mm
sensitivity = 157.08 mm / 1 mm = 157 times (amplification)
one division = 157.08 / 100 = 1.57 mm on the dial, or 0.01 mm of rod travel
Every cent becomes a 1.57-mm line: that's why the comparator "magnifies" the differences
Example 31.4 · Error and Hysteresis from Verification Data

A comparator is checked every 0.2 mm, first moving up and then down. The errors found, in micrometers, are:

Position (mm)00,20,40,60,81,0
rise (µm)012234
slope (µm)234344
difference (µm)222110
maximum error = the larger of the two values = 4 µm
hysteresis = the greatest difference between the descent and ascent at the same point = 2 µm
Maximum error 4 µm, hysteresis 2 µm: compare these values with the manufacturer's specifications
0123450.00.20.40.60.81.0hysteresis2 µmerror in µmmm● rise● descentmaximum error 4 µm
Figure 31.4 · The data from Example 31.4 plotted on a graph: the distance between the two curves represents the hysteresis, and the highest point represents the maximum error. If the instrument were perfect, the two curves would lie on the horizontal axis.

31.5 Measurement Errors

Measurement error is the difference between the measured value and the true value. The true value is unknown, but it can be approximated very closely by measuring a calibrated sample, such as a standard block; in that case, the instrument’s error becomes apparent. Errors are divided into three categories, each of which is treated differently.

Gross errors

These are genuine errors: a misread number, an unchecked zero, a vernier scale read at the wrong mark, a chip between the micrometer’s jaws, a workpiece that’s still hot, or a dimension transcribed backward (25.021 instead of 25.012). There are no statistical rules for these: they are avoided through careful attention, cleanliness, method, and repetition. If a reading is significantly different from the others, investigate the cause before using it.

Systematic Errors

They always repeat identically, with the same sign, under the same conditions: a zero point shifted by one micrometer, a steel gauge measuring an aluminum workpiece at 26 °C (Section 31.10), a tilted dial indicator (Section 31.12). Repeating the measurement is pointless, because each reading is off by the same amount. These errors are identified by comparing the instrument with a standard, and when known, they are corrected.

Random errors

They vary from one reading to the next—sometimes a little more, sometimes a little less—and cannot be predicted: small differences in force, position, temperature, dust, and vibrations. They cannot be corrected individually, but they can be reduced by averaging multiple readings and can be estimated using statistics (Section 31.6).

precise and accurateminor errorsprecise, but not accuratesystematic erroraccurate, not preciserandom errorneither one nor the otherboth
Figure 31.5 · The center of the target is the true value; the red dots are the readings. Precision refers to how close the shots are to each other; accuracy refers to how close their center is to the target.

Two words that are often confused:

  • Accuracy: how closely repeated measurements resemble one another. It depends on random errors. An accurate measurement yields consistent readings;
  • Accuracy: how close the average of the readings is to the true value. It depends primarily on systematic errors. (In the terminology of standards, the closeness of the average to the true value is more precisely called “trueness,” and “accuracy” encompasses both trueness and precision.)

The second point highlighted in the figure is the most dangerous error in the workshop: readings that are all the same may seem reliable, but they are all off by the same amount. Only a sample check can reveal this.

How to Correct Systematic Errors

A sample of known value is measured using the same instrument under the same conditions. The difference between the reading and the sample’s value is the systematic error at that point; the correction is the same number with the sign reversed, and it is added to all subsequent readings.

Systematic Error and Correction
E = xread − xsample C = −E xcorrect = xread + C
E systematic error in mm · xread instrument reading on the sample in mm · xsample certified value of the sample in mm · C correction in mm, to be added to the readings on the workpiece
Example 31.5 · The shifted zero of the micrometer

A 0–25 mm micrometer, placed on a 20.000 mm block, reads 20.003 mm. Immediately afterward, when placed on the workpiece, it reads 18.467 mm.

E = 20.003 − 20.000 = +0.003 mm (the micrometer reads 3 µm more)
C = −0.003 mm
correct dimension = 18.467 + (−0.003) = 18.464 mm
18.464 mm; better yet: adjust the micrometer's zero and double-check it on the block

A correction determined at a single point is reliable only in the immediate vicinity of that point. For an instrument used across the entire measurement range, the complete error curve (as shown in Figure 31.4) is required; this can be found on the calibration certificate.

31.6 Random Errors: Mean and Standard Deviation

When the same measurement is repeated several times, the readings are not all the same. They are expressed as three simple numbers.

Mean, Range, and Standard Deviation
x̄ = x1 + x2 + … + xnn vi = xi − x̄
s = √v1² + v2² + … + vn²n − 1 sx̄ = s√n
x̄ average of the readings · n number of views · vi deviation of reading i from the mean (the sum of the deviations is always zero) · s experimental standard deviation, which indicates how widely the individual readings are scattered · sx̄ standard deviation of the mean, which indicates the degree of uncertainty in the mean. All values are in the same unit as the readings (mm or µm).

Why is s divided by n − 1 rather than by n in the formula? Because the average was calculated from the same readings, which makes them appear slightly more concentrated than they actually are; dividing by n − 1 compensates for this. With many readings, the difference becomes negligible. Scientific calculators have an s (or σn−1) key: that’s the right one.

Example 31.6 · Ten Readings with a Micrometer

The diameter of a pin, all in the same section, is measured ten times with a thousandth-millimeter micrometer. The readings are: 25.012 · 25.009 · 25.014 · 25.011 · 25.015 · 25.010 · 25.013 · 25.012 · 25.014 · 25.010 mm. To simplify the calculations and avoid so many zeros, we work in micrometers above 25.000: 12, 9, 14, 11, 15, 10, 13, 12, 14, 10.

sum = 12 + 9 + 14 + 11 + 15 + 10 + 13 + 12 + 14 + 10 = 120 µm
x̄ = 120 / 10 = 12 µm → average = 25.012 mm
scrap vi: 0, −3, +2, −1, +3, −2, +1, 0, +2, −2 (sum = 0, check is correct)
squares: 0 + 9 + 4 + 1 + 9 + 4 + 1 + 0 + 4 + 4 = 36 µm²
s = √(36 / 9) = √4 = 2.0 µm
sx̄ = 2.0 / √10 = 2.0 / 3.162 = 0.63 µm
average 25.012 mm · each reading has a spread of approximately 2 µm · average uncertainty of only 0.63 µm
0129101112131415average 12micrometers over 25,000 mm−2s−sx̄+s+2s68.3%95.4%Countless readings: the bell
Figure 31.6 · On the left is the histogram of the ten readings from Example 31.6: how many times each value appears. On the right, what happens with a very large number of readings: a bell curve (normal distribution) forms, with 68.3% of the readings within ±s of the mean and 95.4% within ±2s.

The bell curve explains why values are often multiplied by 2 in metrology: approximately 95% of the values fall within ±2s, so a wide interval of ±2s almost always “captures” the correct value. 99.7% fall within ±3s: this is the reasoning we will encounter in statistical process control (Chapter 38).

Example 31.7 · How many readings are needed?

Using the same micrometer (s = 2 µm), we want an average with a standard deviation of no more than 0.5 µm. How many readings do we need to take?

s / √n ≤ 0,5 → √n ≥ 2 / 0,5 = 4 → n ≥ 4² = 16
At least 16 readings. To halve the uncertainty of the average, you need to quadruple the number of readings: beyond a certain point, it’s better to use a higher-quality instrument.

31.7 Measurement Uncertainty

Uncertainty is an interval around the measured value within which, with a high and specified probability, the true value lies. It is not an error to be corrected: it is the doubt that remains after correcting everything that could be corrected. The method for calculating it is described in an international guide (the GUM, also adopted as a UNI CEI standard); here we’ll look at the simplified version that’s sufficient for use in a workshop.

The Two Approaches: Type A and Type B

  • Type A uncertainty: This is calculated using the repeated-measure statistics. It is the standard deviation of the mean, s/√n, as described in paragraph 31.6.
  • Type B uncertainty: This is derived from all other factors: the calibration certificate, the maximum permissible error specified by the manufacturer, resolution, knowledge of the temperature, and experience. When it is known only that the error lies within ±a, with no preference for either value, the half-amplitude is divided by √3 (rectangular distribution). If the certificate specifies an expanded uncertainty U with a factor k, it is divided by k.
Most Common Type B Uncertainties
u = a√3 uris = r2 · √3 = 0,29 · r ucert = Uk
a half-range of the error interval (for a maximum permissible error of ±3 µm, a = 3 µm) · r instrument resolution (the reading may be off by half a resolution in either direction) · U Extended uncertainty stated on the certificate · k certificate coverage factor, usually 2

Resolving Uncertainties: The Sum in Quadrature

The various causes are independent: it is unlikely that they would all be wrong at the same time and in the same direction. For this reason, they are not added directly, but rather by squaring the sum, like the sides of a right triangle: the squares are added, and the square root is taken. The combined uncertainty uc corresponds to one standard deviation, that is, approximately a 68% probability; to obtain approximately a 95% probability, it is multiplied by the coverage factor k = 2, yielding the expanded uncertainty U.

Composite and Extended Uncertainty
uc = √(u1² + u2² + … + um²) U = k · uc result = x̄ ± U
u1, u2, … standard uncertainties of each source, all in the same unit · uc combined standard uncertainty · k coverage factor (k = 2: approximately 95% probability) · U expanded uncertainty · x̄ measured value, already corrected for known systematic errors
Example 31.8 · The uncertainty budget for the pin

Let’s revisit the ten readings from Example 31.6, taken with a 25–50 mm micrometer with a resolution of 0.001 mm. The manufacturer specifies a maximum permissible error of ±3 µm; the parallelism of the measuring faces is guaranteed to be within ±1 µm; the workpiece and the micrometer, both made of steel, may differ in temperature by up to 2 °C.

ReasonTypeCalculationu (µm)u² (µm²)
repeatability of readingsA2,0 / √100,630,400
1 µm resolutionB1 / (2 · √3)0,290,083
micrometer error ±3 µmB3 / √31,733,000
Face parallelism ±1 µmB1 / √30,580,333
temperature: 11.5 × 10⁻⁶ × 25 × 2 = 0.58 µmB0,575 / √30,330,110
sum of squares3,927
uc = √3.927 = 1.98 µm
U = 2 × 1.98 = 3.96 µm, rounded up to 4 µm = 0.004 mm
Ø = 25.012 ± 0.004 mm (k = 2, approximately 95% confidence level)

The factor that weighs most heavily is the instrument’s error: it alone accounts for three-quarters of the total. Taking 100 readings instead of 10 would make almost no difference; to improve the result, we would need a better micrometer or a comparative measurement using blocks.

How to Write a Result

  • "uncertainty" is spelled with one or, at most, two significant digits, rounded by excess (it's better to be a little more uncertain than the opposite);
  • The value is rounded to the same decimal place as the uncertainty: it makes no sense to write 25.01237 if the uncertainty is 4 thousandths;
  • The unit and the coverage factor are always specified: 25.012 ± 0.004 mm (k = 2).
Example 31.9 · Rounding Properly

A spreadsheet returns a value of 12.34567 mm and an expanded uncertainty of 0.01231 mm.

Uncertainty with two significant figures, rounded up: 0.01231 → 0.013 mm
value with the same number of decimal places: 12.34567 → 12.346 mm
12.346 ± 0.013 mm (k = 2). Incorrect: “12.34567 ± 0.01231” (unnecessary digits) and “12.35 ± 0.013” (value truncated too much)

31.8 The Rule of Tithing and the Choice of Instrument

Before taking a measurement, you should ask yourself whether the instrument is suitable for the tolerance. The most widely used rule of thumb is the one-tenth rule: the measurement uncertainty (and therefore also the instrument’s resolution) must not exceed one-tenth of the tolerance to be checked.

Rule of Tenths
U ≤ T10 r ≤ T10
U: expanded measurement uncertainty · r: instrument resolution · T: total tolerance (difference between the upper and lower limits: for 50 ± 0.05, T = 0.1 mm), all in mm

If this rule cannot be followed (for example, for tolerances of a few micrometers), a ratio of 1/5 to 1/4 is acceptable; however, in that case, the uncertainty significantly reduces the useful tolerance range, as shown in paragraph 31.9. The table provides approximate values for instruments in good condition, used by an experienced operator in a typical workshop environment; the actual values can be found on the certificates.

ToolResolution (mm)Typical uncertainty (mm)Minimum tolerance using the one-tenth rule (mm)
Vernier Caliper0,02 o 0,05from 0.03 to 0.05from 0.3 to 0.5
Digital Caliper0,01from 0.02 to 0.03from 0.2 to 0.3
Centesimal Micrometer0,01from 0.005 to 0.01from 0.05 to 0.1
Millesimal Micrometer0,001from 0.002 to 0.004from 0.02 to 0.04
Centesimal comparator (for comparison)0,01from 0.005 to 0.01from 0.05 to 0.1
Thousandth-of-an-inch comparator on blocks0,001from 0.001 to 0.002from 0.01 to 0.02
Electronic comparator in the metrology lab0,0001from 0.0003 to 0.0005from 0.003 to 0.005
Coordinate measuring machine (Chapter 37)0,0001from 0.002 to 0.005from 0.02 to 0.05

Approximate and rounded values: The actual uncertainty depends on the model, dimensions, method, and environment.

Example 31.10 · What tool is used for Ø40 h7?

A shaft with a diameter of Ø40 and a height of h7 has a tolerance of IT7 = 25 µm (Chapter 14): from 39.975 to 40.000 mm.

U ≤ T / 10 = 0.025 / 10 = 0.0025 mm = 2.5 µm
digital caliper (U ≈ 0.02): no · thousandth-millimeter micrometer (U from 0.002 to 0.004): barely acceptable, only if in good condition
Millesimal comparator zeroed on 40-mm blocks (U ≈ 0.001 to 0.002): yes
In production: go/no-go gauge or comparator on blocks; the thousandth-of-an-inch micrometer is used only for occasional checks
Example 31.11 · What is the range of the digital caliper?

According to its certificate, a digital caliper has an uncertainty of 0.02 mm. What dimensions can it measure while adhering to the rule of tenths?

T ≥ 10 · U = 10 · 0.02 = 0.2 mm
50 ± 0.1 (T = 0.2): yes, barely · 50 ± 0.05 (T = 0.1): no · 50 ± 0.2 (T = 0.4): yes, by a wide margin
For tolerances tighter than ± one-tenth, a micrometer is used

31.9 Determining Whether a Workpiece Meets Specifications: The Zones in ISO 14253-1

A workpiece measures 25.018 mm; the tolerance is up to 25.020 mm; the uncertainty is 0.004 mm. Is it acceptable? It’s hard to say: the true value could be 25.014 (good) or 25.022 (out of tolerance). The ISO 14253-1 standard specifies how to make this determination while taking uncertainty into account, and divides the scale into zones.

  • Conformance zone: the narrow tolerance of uncertainty U on either side. A measurement within this zone proves that the workpiece is in conformance.
  • Nonconforming zone: outside the expanded tolerance of U on either side. A measurement here indicates that the workpiece is nonconforming.
  • Areas of uncertainty (doubtful): the two 2U-wide bands straddling each limit. Here, the measurement does not prove either one or the other.
non-compliantdoubtfulcompliantdoubtfulnon-complianttolerance: 25 ± 0.020 mmmeasurement 25.012 ± 0.004measurement 25.018 ± 0.004measurement 25.026 ± 0.004−40−24−20−160+16+20+24+40Deviation from 25,000 mm in µm · U = 4 µm
Figure 31.7 · Tolerance of 25 ± 0.020 mm measured with U = 0.004 mm. The green zone (from 24.984 to 25.016) indicates compliance, the red zones indicate non-compliance, and the yellow zones are uncertain. The bars show three measurements along with their ±U.
Zone Boundaries (ISO 14253-1)
Compliance: from LSL + U to USL − U; Non-compliance: below LSL − U or above USL + U
LSL: lower specification limit (minimum dimension) · USL: upper specification limit (maximum dimension) · U: expanded measurement uncertainty, all in mm

The rule is based on a common-sense principle: the party that must prove something bears the burden of uncertainty. The supplier who wants to prove that the workpiece is acceptable uses the narrow zone; the customer who wants to prove that the workpiece should be rejected must find it outside the expanded zone. In borderline zones, the decision is based on an agreement between the parties, or the part is remeasured using a more precise method (smaller U, narrower borderline zones).

Example 31.12 · Three pins, three answers

Tolerance 25 ± 0.020 mm (LSL = 24.980, USL = 25.020), uncertainty U = 0.004 mm. Three pins were measured: 25.012, 25.018, and 25.026 mm.

tolerance range: from 24.980 + 0.004 = 24.984 to 25.020 − 0.004 = 25.016 mm
Non-conformity: below 24.980 − 0.004 = 24.976 or above 25.020 + 0.004 = 25.024 mm
25,012: within the green zone → compliant
25.018: between 25.016 and 25.024 → uncertain; cannot be classified as either acceptable or rejected
25,026: more than 25,024 → non-compliant
The second pin is within tolerance on the display, but the supplier cannot prove it given this uncertainty

Note that here U = 0.004 mm with a tolerance of T = 0.040 mm: this is exactly the one-tenth rule, yet the compliance zone is already 20% narrower than specified in the drawing (32 µm instead of 40). With a 1/4 ratio, half of the tolerance would be lost.

31.10 Reference Temperature and Expansion

All materials expand when heated (Chapter 14, Hot Assembly). For this reason, a dimension is meaningful only if the temperature at which it applies is specified. ISO 1 sets the reference temperature at 20 °C: the dimensions in the drawing, the tolerances, and the measurements of blocks and tools are all valid at 20 °C.

Linear expansion
ΔL = α · L · Δt Δt = t − 20 °C
ΔL change in length in mm · α coefficient of linear expansion in 1/°C · L length in mm · Δt difference between the temperature of the workpiece and 20 °C, in °C (or K)
Materialα (10⁻⁶ /°C)Materialα (10⁻⁶ /°C)
Carbon steel, steel blocks11,5Aluminum and Alloys23
Gray cast iron10Brass19
Austenitic stainless steel16Copper17
Titanium8,6Granite (countertops)6 to 8
Hard metal5 to 6Engineering plastics (POM, PA)from 80 to 120

Average values are rounded and approximate; actual alloys may vary.

The workpiece and the instrument expand together

If the workpiece and the measuring tool are made of the same material and are at the same temperature, they expand by the same percentage, and the reading remains accurate even at temperatures other than 20 °C: this is why micrometers, calipers, and gauge blocks are made of steel, just like most workpieces. Problems arise when:

  • the workpiece is made of a different material (aluminum, brass, plastic) and the measurement is not taken at 20 °C;
  • The workpiece and the tool are at different temperatures: the workpiece has just come out of the machine, while the tool has been warmed by hands or the sun.

When the temperatures are the same but the materials are different, the length at 20 °C is calculated as follows:

Correction of a measurement taken at temperature t
L20 = Lletta · 1 + αs · (t − 20)¹ + αp · (t − 20) ≈ Lletta + L · (αs − αp) · (t − 20)
L20 Length of the workpiece at 20 °C in mm · Lread instrument reading in °C in mm · αs instrument factor (or factor of the sample used for zero adjustment) · αp workpiece coefficient, in 1/°C · t The common temperature of the workpiece and the tool in °C. An approximate value is always sufficient in the workshop.
20 °Cworkpiece of aluminumsteel ruler26 °Caluminum workpiece+13.8 µm+27.6 µmLength 200 mm at 20 °C; elongations greatly exaggerated in the drawing.
Figure 31.8 · At 20 °C, the aluminum workpiece and the steel ruler align. At 26 °C, the steel ruler has elongated by 13.8 µm, and the aluminum workpiece by 27.6 µm: the instrument detects only the difference and measures the workpiece as 13.8 µm longer than it is at 20 °C.
Example 31.13 · Aluminum measured at 26 °C

In a room at 26 °C, a workpiece of aluminum left for a long time next to the instrument (and therefore at the same temperature) is measured with a steel micrometer: 200.020 mm. What is its measurement at 20 °C? (αs = 11.5 × 10⁻⁶, αp = 23 × 10⁻⁶ /°C)

t − 20 = 26 − 20 = 6 °C
correction = 200 · (11.5 − 23) · 10⁻⁶ · 6 = 200 · (−11.5 · 10⁻⁶) · 6 = −0.0138 mm
L20 = 200.020 − 0.0138 = 200.0062 mm
using the exact formula: 200.020 × (1 + 0.000069) / (1 + 0.000138) = 200.0062 mm
200.006 mm at 20 °C: without correction, they would have “given away” 14 µm per workpiece
Example 31.14 · The workpiece just removed from the machine

A Ø100 steel shaft comes off the lathe at 30 °C and is immediately measured with a micrometer that has been stored in a cabinet at 20 °C.

ΔL = 11.5 × 10⁻⁶ × 100 × (30 − 20) = 0.0115 mm
The shaft appears to be 11.5 µm larger: an h6 (tolerance 22 µm) loses half of its tolerance. When it cools, the “properly turned” workpiece ends up undersized

Stabilization time and hand temperature

A workpiece brought into the metrology lab must remain on the work surface until it reaches room temperature. The time required depends on the mass and material: from a few tens of minutes for a small piece of aluminum to many hours for a steel block weighing tens of kilograms. A reliable rule of thumb is to place it on a cast-iron or granite surface (which acts as a thermal mass) and take a measurement when two readings taken 15 minutes apart yield the same value.

The operator’s body is also a source of heat: the hands are over 30 °C. Holding a block, a zeroing rod, or the jaw of a large micrometer stretches them by a few micrometers in just a few minutes. This is why micrometers have insulating plates on the jaw, collets are used to handle blocks, and small workpieces are placed gently rather than held tightly.

Example 31.15 · The Notepad Held in the Hand

A 100-mm steel block, when held in the hand for a couple of minutes, heats up by 2 °C.

ΔL = 11.5 × 10⁻⁶ × 100 × 2 = 0.0023 mm = 2.3 µm
2.3 µm: more than the entire tolerance of a Class 0 block of that size. The block is placed and left to settle.

31.11 Abbe's Error

Ernst Abbe, a 19th-century German physicist, formulated the principle that bears his name: the instrument’s scale must lie on the same line as the dimension to be measured. If there is a distance h between the scale line and the measurement line, any slight tilt of the cursor (due to clearances, wear, or the operator’s force) results in an error proportional to h.

Abbe's Error
e = h · tan θ
and error in mm · h distance between the scale line and the measuring line (the “Abbe arm”) in mm · θ angle of inclination of the slider or probe. For small angles, tan θ is almost equal to the clearances / lengths of the guide.
scale linefixed spoutslider with clearanceworkpiecehθe = h · tan θThe scale reads the positionof the cursor, not that of thepoint of contact with the workpiece
Figure 31.9 · In a vernier caliper, the scale is on the rule and the workpiece is positioned lower, at a distance h. Due to clearance in the rule, the slide tilts by an angle θ (greatly exaggerated in the drawing): the jaw touches the workpiece further back by e than where the scale indicates it to be.
Example 31.16 · How much error does the caliper introduce when measuring at the tip?

The slide of a caliper is 40 mm long and has a clearance of 0.02 mm at the scale. A workpiece is measured at the end of the jaws, at h = 30 mm from the scale, and then near the scale, at h = 10 mm.

tan θ = 0.02 / 40 = 0.0005 (θ = 0.029°)
at the tips of the beaks: e = 30 × 0.0005 = 0.015 mm
near the line: e = 10 · 0.0005 = 0.005 mm
One and a half cents versus half a cent: when using a caliper, always measure as close to the line as possible without clamping too tightly

The micrometer, on the other hand, follows Abbe’s principle: the micrometer screw, which serves as its scale, lies on the same axis as the anvil and the rod. An inclination of the rod no longer results in an error proportional to the inclination, but only in a very small cosine error (Section 31.12). This is one of the reasons why the micrometer is more precise than the vernier caliper, in addition to its higher resolution.

workpieceMeasuring axis = scale axis: h = 0anvilrodbushdrumclutchbow
Figure 31.10 · In an anvil micrometer, the workpiece, the rod, and the micrometer screw are aligned (red dashed line): the Abbe arm is zero.
Example 31.17 · The Same Angle on the Micrometer

Suppose that the stem of a micrometer, which is 25 mm long, is tilted by the same angle as the gauge (tan θ = 0.0005). The error is now given by cosine: L · (1 − cos θ).

cos θ = cos 0.029° = 0.999999875
e = 25 · (1 − 0.999999875) = 0.000003 mm = 3 nanometers
Negligible: Applying Abbe transforms a first-order error into a second-order error

12/31 Parallax, Measurement Force, Cosine, and Positioning

Parallax error

In analog instruments, the indicator (the pointer, the edge of the vernier scale, or the line on the drum) does not lie exactly on the plane of the scale, but slightly above it, at a distance g. If the eye is not looking perpendicular to the scale, the indicator will appear to point to a different mark. Solution: Always read with your eye aligned perpendicular to the pointer; some comparators have a small mirror on the dial, and the reading is correct when the pointer covers its reflected image. Digital instruments do not have parallax.

Table of Contentseye above the indexside viewgβe = g · tan βcorrect readingincorrect reading
Figure 31.11 · The pointer is at a height g above the scale. When viewed from the side at an angle β, the reading shifts by e = g · tan β.
Example 31.18 · Parallax on a comparator and a caliper
comparator: pointer 1 mm above the dial, eye tilted at 15°: e = 1 · tan 15° = 0.268 mm on the dial
One division is 1.571 mm long and represents 0.01 mm: 0.268 / 1.571 × 0.01 = 0.0017 mm
caliper: 0.1 mm vernier above the scale, eyepiece tilted at 30°: e = 0.1 · tan 30° = 0.058 mm
On the comparator, 1.7 µm (a small amount); on the caliper, nearly 6 centesimi: three vernier divisions of 0.02

Measuring Force and Deformation

To take a measurement, you must touch the workpiece, and touching means applying pressure. The measuring force elastically compresses the contact point, bends the micrometer’s arc, and, on thin workpieces, deforms the workpiece itself. To keep it constant, the micrometer has a clutch (or ratchet) on the drum, which always slips at the same force, usually between 5 and 10 N; dial indicators have a spring that pushes the stem with 1 or 2 N. The rule is: turn only the clutch, never the drum while holding it between your fingers.

When a spherical tip presses against a flat surface, the contact is not a point but a small elastic indentation, and the tip “sinks” by a distance δ, which is estimated using Hertz’s theory:

Elastic compression of a sphere on a plane (Hertz, simplified form)
δ = ∛9 · F²16 · R · E*² 1E* = 1 − ν1²E1 + 1 − ν2²E2
δ elastic deflection in mm · F measured force in N · R nose radius in mm · E* equivalent elastic module in N/mm² · E1, E2: elastic modules of the tip and the workpiece (steel 210,000, aluminum 70,000 N/mm²) · ν: Poisson’s ratio (steel 0.30, aluminum 0.33)
Example 31.19 · How far does the dial indicator's probe sink?

Spherical steel tip, Ø3 (R = 1.5 mm), measuring force 1.5 N, on a steel workpiece and then on an aluminum workpiece.

steel on steel: 1/E* = 2 · (1 − 0.09) / 210,000 → E* = 115,385 N/mm²
δ = ∛(9 · 1.5² / (16 · 1.5 · 115,385²)) = ∛(20.25 / 3.195 × 10¹¹) = ∛(6.34 × 10⁻¹¹) = 0.00040 mm
steel on aluminum: 1/E* = 0.91 / 210,000 + 0.8911 / 70,000 → E* = 58,605 N/mm² → δ = 0.00063 mm
0.40 µm on steel, 0.63 µm on aluminum

These are small numbers, but when compared to steel blocks, the difference (0.23 µm) on an aluminum workpiece is no longer negligible. With double the force, the compression increases by a factor of 22/3 = 1.59. On thin-walled tubes, spring rings, and plastics, the deformation of the workpiece is much greater: in these cases, reduced forces or non-contact tools are used (Chapter 37).

Cosine Error of the Comparator

The stylus of a dial indicator must move in the same direction as the displacement being measured—that is, perpendicular to the surface. If it is tilted by an angle α, the stylus must travel a greater distance—d / cos α—to track a displacement d of the surface; the reading is greater than the actual value. The same thing happens with a lever-type comparator (dial indicator) when the arm is not parallel to the surface.

Cosine Error
true value = reading · cos α error = reading · (1 − cos α)
α: the angle between the rod (or the working plane of the lever arm) and the direction of measurement · reading and error in the same unit
dperpendicular rodreading = ddαrod inclined by αreading = d / cos αtrue value = reading · cos α
Figure 31.12 · The surface rises by d (dashed line). With the rod perpendicular, the reading is d; with the rod inclined by α, the rod moves by d / cos α, and the reading is too high.
Example 31.20 · Misaligned comparator

When checking the eccentricity of a shaft, the dial indicator is tilted 15° and reads 0.250 mm.

cos 15° = 0.9659
true value = 0.250 × 0.9659 = 0.2415 mm
error = 0.250 · (1 − 0.9659) = 0.0085 mm
0.2415 mm: the error is nearly one hundredth. Below 5°, the error is less than 0.4% of the reading

Positioning Errors

The instrument must touch the workpiece at the correct points. An outside diameter measured with the micrometer tilted appears larger than it actually is; a hole measured with the tips offset from the center yields a chord that is shorter than the diameter. For this reason, the instrument is oscillated within the hole to find the maximum value, while on a shaft or between two flat surfaces, the minimum value is sought. Bore gauges (Chapter 32) have centering supports specifically for this purpose.

D (in the center)string < DandThe tips of the caliper or bore gaugeshifted by e from the center measurea chord that is progressively shorter than D.chord = 2 · √(R² − e²)For this reason, the maximum is sought in the hole,and the minimum on the shaft and between two planes.
Figure 31.13 · As the tips move toward and away from the center, they measure a chord (in red) instead of the diameter (in green). This displacement is greatly exaggerated.
Example 31.21 · The hole was measured off-center

A Ø50 hole (R = 25 mm) is measured with the probes offset by e = 1 mm from the center.

cord = 2 · √(25² − 1²) = 2 · √624 = 2 · 24.980 = 49.960 mm
error = 50.000 − 49.960 = 0.040 mm
Just 4 cents less for a single millimeter off-center: the hole seems too tight even when it isn't

31.13 Cleaning, reference surface, and metrology room

Cleaning and Deburring

A strand of hair is about 0.05 mm thick, a film of oil is a few micrometers thick, and a speck of dust is 1 to 10 µm in size. These quantities are all comparable to the tolerances. Before taking any measurement:

  • Clean the measuring faces of the instrument (by placing a sheet of clean paper between the faces of the micrometer and pulling it out smoothly) and the surfaces of the workpiece;
  • Remove the coolant and chips, including those inside the holes and threads;
  • Remove any burrs: a burr on an edge can lift the workpiece off the surface or keep the jaws of the micrometer open, and the measurement will then reflect the size of the burr. Workpieces with burrs cannot be measured.

The reference plane

The reference surface is the surface on which the workpieces, the blocks, and the dial indicator stand rest. It is made of two materials:

  • granite (actually often diabase or black gabbro): it does not rust, is not magnetized, expands very little, remains stable over time, and, when struck, chips without raising a bump. It is the most commonly used material in metrology laboratories;
  • Stabilized and scraped cast iron: more impact-resistant, suitable for workshop use and marking, but it must be protected from rust.
granite countertopsmall blocksworkpiececomparatoron a standthree supports: the surface does not warp
Figure 31.14 · Comparative measurement on the reference surface: Zero the dial indicator on the blocks and move it to the workpiece by sliding the base across the surface. The surface rests on three points (shown in red).

The floors are constructed of flatness classes (00, 0, 1, 2, 3 according to DIN 876, or the equivalent classes in ISO 8512): Classes 00 and 0 have deviations of only a few micrometers even on surfaces one meter in length and are intended for metrology laboratories; Classes 1 and 2 are suitable for workshop inspections; Class 3 is for marking out. The exact value depends on the dimensions and is specified on the surface’s certificate. The surface rests on three supports Adjustable, because three points always define a plane and the slab does not warp; it should be checked periodically because it wears down more in the area where work is always being done.

The Metrology Lab

The most precise measurements are taken in a dedicated room—the metrology lab—under controlled conditions. Approximate values:

ConditionTypical valueWhy
Temperature20 °C ± 1 °C (± 0.5 °C for more precise measurements)ISO 1 reference, controlled expansion
Change over timeless than 1 °C per hourThe workpieces and instruments keep up in time
The Difference Between High and Lowless than 1 °C per meterThe workpiece must not have any part that is hotter
Relative humidity40 to 60 percentNo rust on the instruments, no condensation
Vibrationslow; floors and machines on isolated basesMicrometer readings must not waver
Cleanliness and Lightfiltered dust, even lighting without direct sunlightdust on their faces, parallax, the heat from the lamps

Even those who work in a workshop can follow these guidelines: keep tools away from windows, radiators, and engines; make sure workpieces are securely positioned on the working plane; and do not leave the micrometer on the machine in direct sunlight.

31.14 Calibration, Certificates, and Instrument Management

Three words not to confuse:

  • calibration: the comparison of an instrument with a traceable standard to determine its errors and uncertainty. Calibration does not “adjust” anything; it measures and documents;
  • adjustment (or calibration): the process of bringing the instrument closer to the correct setting, such as resetting the micrometer to zero using the adjustment screw. After an adjustment, the calibration must be repeated;
  • verification (or metrological confirmation): the comparison of calibration results with the limits established by the company, which leads to a decision: whether the instrument should be used, used with restrictions, or taken out of service.

The Calibration Certificate

A reliable certificate must include at least the following: the instrument’s identification (brand, model, serial number, or internal code), the date, environmental conditions, the samples used along with their traceability, the method, the results item by item (sample value, reading, error), and the expanded uncertainty with its coverage factor. Certificates from accredited centers bear the ACCREDIA mark and the center’s LAT number. The certificate does not state that the “instrument is good”; this is determined by the company by comparing errors and uncertainty with its own standards, using the same reasoning as in ISO 14253-1.

Example 31.22 · Does the micrometer pass the test?

The company accepts a maximum error of ±2 µm for micrometers ranging from 0 to 25 mm. The certificate specifies, with U = 0.6 µm (k = 2), an error of +1.2 µm at 10 mm and +1.7 µm at 20 mm.

at 10 mm: 1.2 + 0.6 = 1.8 µm ≤ 2 µm → compliant
at 20 mm: 1.7 + 0.6 = 2.3 µm > 2 µm, but 1.7 < 2 µm → uncertain range
The instrument cannot be declared compliant across its entire range: it must be adjusted and recalibrated, or used with a limited use label.

Calibration Intervals

There is no legally mandated calibration interval that applies to everyone: the company determines the interval based on how often the instrument is used, the environment, the required accuracy, and the history of previous calibrations (if the instrument is well within the limits each time, the interval can be extended; if it is out of range, the interval is shortened, and the workpieces measured in the meantime are evaluated). Indicative starting values:

ToolApproximate range
Calipers and Micrometers for Everyday Use in Manufacturing6 to 12 months
Comparators and Bore Gauges12 months
Pass/Fail Gauges (plugs, forks, rings)3 to 12 months, depending on the number of workpieces inspected
Workshop Reference Blocks1 to 2 years
Sample blocks from the metrology lab2 to 3 years
Reference planes1 to 3 years

Between calibrations, interim checks are performed: the micrometer is zeroed daily or at the start of each shift, and a check is performed on a calibration block if the instrument falls or gives unusual readings.

Status Labels

Each instrument has a unique code (engraved or on a tag) and a label that shows at a glance whether it can be used. The most common colors are as follows:

CALIBRATEDCode: MIC-014Next calibration: 03/2027LIMITED USECode: CAL-007Valid only from 0 to 100 mmOUT OF SERVICECode: COM-003Do not use: needs calibration
Figure 31.15 · Green: Calibrated and usable until the date indicated. Yellow: Usable only within the specified limits. Red: Do not use. Instruments used only for indication, without a measured value, are labeled “not subject to calibration.”

An instrument that has expired, is unlabeled, or has been dropped must not be used to measure workpieces until it has been rechecked. If calibration reveals that it was out of tolerance, you must determine which workpieces were measured with it since the last successful calibration: this is why the instrument log must be kept up to date (Chapter 38).

Management and Conservation

To DoDon't Do This
Put each instrument back in its case, cleanLeave it on the machine among the chips and coolant
Protect the steel parts with a thin layer of light oil or an anti-corrosion productTouching the measuring surfaces with sweaty fingers
Store the micrometer with the jaws slightly openClose it tightly: the faces get damaged and the bow remains taut
Slowly bring the instrument closer and use the clutchUsing a caliper as a hammer, compass, or wrench
Remove the batteries from digital devices that have not been used for a long timeMeasuring a rotating workpiece or one that is still hot
Report a fallen tool immediatelyPut it back in the drawer, hoping it hasn't gotten ruined
Keep reference samples separate from test samplesDo not bring steel tools near magnets or magnetic spindles
🖥️
Try it out in CNC Code. In the simulator, you can machine a workpiece, measure it, and adjust the tool offset—it’s the same principle as in this chapter. A measurement error of one cent becomes an incorrect offset of one cent on all subsequent workpieces.

31.15 The Formulas in This Chapter

What Are You Looking For?FormulaUnits
Comparison Measuredimension = sample + readingmm
Systematic Error and CorrectionE = xread − xsample · C = −Emm
Mediax̄ = (x1 + … + xn) / nmm
Standard Deviations = √(Σ vi² / (n − 1))mm or µm
Standard Deviation of the Mean (Type A)sx̄ = s / √nmm or µm
Rectangular Type Bu = a / √3mm or µm
Type B from the resolutionu = r / (2 · √3) = 0,29 · rmm or µm
Type B as indicated on the certificateu = U / kmm or µm
Composite Uncertaintyuc = √(u1² + u2² + …)mm or µm
Widespread UncertaintyU = k · uc (k = 2)mm or µm
Rule of TenthsU ≤ T / 10 · r ≤ T / 10mm
Compliance Zone (ISO 14253-1)from LSL + U to USL − Umm
Non-Compliance Zonebelow LSL − U or above USL + Umm
DilationΔL = α · L · (t − 20)mm
Measurement taken at 20 °CL20 ≈ Lletta + L · (αs − αp) · (t − 20)mm
Abbe's Errore = h · tan θmm
Parallax errore = g · tan βmm
Flattening a sphere on a planeδ = ∛(9 · F² / (16 · R · E*²))mm
Cosine Errortrue = reading · cos α · error = reading · (1 − cos α)mm
Chord of an off-center holechord = 2 · √(R² − e²)mm

31.16 Exercises

Try to solve them on your own, using a calculator, before looking at the solution.

Exercise 1. A comparator is zeroed on a 30.000 mm block. On two workpieces, it reads +0.018 mm and −0.006 mm. What are the dimensions of the workpieces?
Solution

30.000 + 0.018 = 30.018 mm; 30.000 − 0.006 = 29.994 mm.

Exercise 2. The circumference of a pulley, as measured with a tape measure, is 2513.3 mm. What is its diameter?
Solution

D = 2513.3 / 3.1416 = 800.0 mm.

Exercise 3. Five measurements of the same diameter: 48.012, 48.015, 48.010, 48.013, and 48.016 mm. Calculate the mean, standard deviation, and standard deviation of the mean.
Solution

In µm, above 48,000: 12, 15, 10, 13, 16; sum 66, average 13.2 µm → 48.0132 mm. Deviations: −1.2; +1.8; −3.2; −0.2; +2.8; squares: 1.44 + 3.24 + 10.24 + 0.04 + 7.84 = 22.8 µm². s = √(22.8 / 4) = √5.7 = 2.39 µm; s / √5 = 2.39 / 2.236 = 1.07 µm.

Exercise 4. An uncertainty budget has three components: 1.2 µm, 0.8 µm, and 0.5 µm. Calculate the combined uncertainty and the expanded uncertainty with k = 2.
Solution

uc = √(1.44 + 0.64 + 0.25) = √2.33 = 1.53 µm; U = 2 × 1.53 = 3.05 µm, rounded up to 3.1 µm.

Exercise 5. A Ø60 H7 hole has a tolerance of 30 µm. What is the maximum uncertainty allowed under the one-tenth rule? Is a digital caliper acceptable?
Solution

U ≤ 0.030 / 10 = 0.003 mm = 3 µm. A digital caliper (U ≈ 0.02 mm) won't work: you need a thousandth-millimeter bore gauge zeroed to a reference ring, or a pass/fail plug gauge.

Exercise 6. Tolerance: 50 ± 0.05 mm; uncertainty U = 0.008 mm. Classify the measurements 50.030, 50.046, and 49.940 mm according to ISO 14253-1.
Solution

Compliant from 49.958 to 50.042; non-compliance below 49.942 or above 50.058. 50.030: compliant. 50.046: questionable (between 50.042 and 50.058). 49.940: non-compliant (below 49.942).

Exercise 7. A workpiece of brass (α = 19 · 10⁻⁶ /°C) measured at 25 °C with a steel micrometer (α = 11.5 · 10⁻⁶ /°C) at the same temperature gives a reading of 150.012 mm. What is its length at 20 °C?
Solution

Correction = 150 · (11.5 − 19) · 10⁻⁶ · 5 = −0.0056 mm; L20 = 150.012 − 0.0056 = 150.0064 mm.

Exercise 8. On a vernier caliper, the workpiece is 25 mm from the scale, and the non-moving jaw is tilted by 0.03°. What is the Abbe error?
Solution

e = 25 · tan 0.03° = 25 · 0.000524 = 0.0131 mm, approximately 13 µm.

Exercise 9. A lever comparator with its arm inclined at 20° to the surface reads 0.050 mm. What is the true value?
Solution

true = 0.050 · cos 20° = 0.050 · 0.9397 = 0.0470 mm.

Exercise 10. A hole with a diameter of 80 is measured with the probes positioned 0.5 mm off-center. What is the reading, and by how much is it incorrect?
Solution

cord = 2 · √(40² − 0.5²) = 2 · √1599.75 = 79.9937 mm; error 80 − 79.9937 = 0.0063 mm, approximately 6 µm.

Exercise 11. What is the standard uncertainty due to the resolution of a 0.01 mm digital caliper?
Solution

u = 0.01 / (2 · √3) = 0.01 / 3.464 = 0.0029 mm.

Exercise 12. In Example 31.19, the tip penetrates 0.40 µm into the steel with a force of 1.5 N. How far does it penetrate with a force of 3 N?
Solution

δ increases as F²/³: 0.40 · 22/3 = 0.40 · 1.587 = 0.63 µm.

Exercise 13. Given a standard deviation of s = 3 µm, we want to obtain an average with a Type A uncertainty of no more than 1 µm. How many readings are needed?
Solution

√n ≥ 3 / 1 = 3 → n ≥ 9 readings.