Chapter 36

Roughness Measurement

In Chapter 16, we learned how to read and write roughness on drawings: Ra, Rz, the classes, and the symbols. Here, we’re switching sides of the bench: how to actually measure it. We’ll see when it’s enough to run your fingernail over a reference sample and when you need a roughness tester, how the instrument is constructed, what the filter does, how to choose the cutoff, where and in which direction to place the probe, how to read the measurement ratio, and how to determine if the workpiece is acceptable. The chapter concludes with a discussion of the most common errors, optical instruments, and calibration.

36.1 Why Is Roughness Measured?

Roughness is not merely an aesthetic consideration. It determines whether a seal ring leaks oil, whether a guide wears out in a month or in ten years, whether a shaft breaks due to fatigue starting from the bottom of a groove, whether a interference fit holds, and whether paint adheres. The recommended roughness values for various functions are listed in the table in Section 16.10; here, we are concerned with the next step: verifying that the workpiece coming off the machine meets these specifications.

Measuring surface roughness is useful not only for inspectors but also for those who actually do the work. An Ra value that increases from workpiece to workpiece is the first sign of a worn insert, a worn cutting edge, or a spindle that is starting to vibrate. Those who monitor roughness can spot problems before they result in scrap.

When a comparison Is Enough, and When a Tool Is Needed

Not all surfaces require a roughness tester. For roughing, for a surface that doesn’t come into contact with anything, or for a quick check right at the machine, it’s enough to compare the workpiece with a sample—by sight and touch. The instrument becomes essential when precision really matters.

SituationAppropriate method
Non-functional surfaces, Ra 6.3 µm or greatervisual and tactile comparison with the samples
In-machine inspection during the production runcomparison, plus a portable roughness tester every now and then
Functional surfaces with a Ra of 1.6 µm or lessroughness meter
Seats, bearing seats, surfaces marked with “max” or with Rz, Rmaxroughness meter, with a cutoff point and carefully selected points
Testing, first workpiece, dispute with the customercalibrated roughness meter, with stored measurement ratio
Very smooth surfaces (Ra below 0.05 µm), soft materials, 3D analysisnoncontact instruments (paragraph 36.13)

Guideline table prepared by CNC Simulator Italia: Each company establishes its own rules in its control plan (Chapter 38).

36.2 Comparative roughness samples

A comparative sample (known in the workshop as a “roughness plate”) is a plate with a series of surfaces of known roughness, produced by a specific machining process. The most common series reproduce turning, peripheral milling, face milling, and flat and cylindrical grinding; there are also series for lapping, EDM, and sandblasting. The surfaces are usually reproduced by electroplating (with nickel) from a machined and measured original, so all the plates in a series are identical.

TurningRa 0.4Ra 0.8Ra 1.6Ra 3.2Ra 6.3Ra 12.5MillingendRa 0.8Ra 1.6Ra 3.2Ra 6.3Ra 12.5Ra 25GrindingFlatRa 0.05Ra 0.1Ra 0.2Ra 0.4Ra 0.8Ra 1.6
Figure 36.1 · A comparison plate with three sets of marks. In turning, the grooves widen with the feed; in face milling, the arcs left by the milling cutter’s path remain; in grinding, the lines are dense and irregular. Approximate values in µm.

How to Use Them

  • Choose a series that matches the machining operations of the workpiece: comparing a turned workpiece with a ground sample makes no sense, because the eye and the finger primarily perceive the shape of the grooves;
  • Clean the workpiece and place it next to the sample, with the grooves facing the same direction, under the same light—preferably at a glancing angle;
  • First, look at it with the naked eye, then with a magnifying glass that magnifies 5 or 10 times;
  • Run your fingernail (or fingertip) over the workpiece and the sample, always perpendicular to the grooves and at the same speed: the fingernail vibrates more on rough surfaces;
  • Find the two samples that “enclose” the workpiece and write the result as a range: “between Ra 0.8 and Ra 1.6.”

The Limits

The comparison is subjective. With a little practice, you can confidently distinguish a step in the scale (a factor of 2 in Ra); distinguishing finer steps is not realistic. The material can be misleading: a workpiece of polished aluminum feels smoother than a nickel sample with the same Ra value, while cast iron feels rougher. Below an Ra of 0.4 µm, the sense of touch can no longer distinguish well. This comparison does not provide Rz or Rmax values and cannot detect a single deep scratch. Finally, the samples get scratched and dirty: they should be kept in their case and replaced when they show signs of wear. For these reasons, the comparison is an excellent production control, but it is not suitable as a test for a functional surface.

✋
Train your finger. A good exercise for beginners: measure five different workpieces with the roughness tester, write the values on a covered sheet of paper, and then try to put them in order using your fingernail and the gauge plate. After a few weeks, the error almost always falls within one class.

36.3 The stylus roughness tester

The most widely used instrument is the stylus-type roughness tester (also known as a contact or probe-type roughness tester). The principle is simple: an extremely fine tip is dragged across the surface at a constant speed, and a transducer records how much it rises and falls. The signal is amplified, converted into numerical values, filtered, and transformed into the parameters described in Chapter 16.

workpiecetranslation unitmotor, guide, encoderprobe body with transducersliderdiamond tipradius 2 or 5 µm, cone 60° or 90°probe travelRa 0.82filter and calculationElectronicsand display
Figure 36.2 · Schematic of a portable roughness tester with a glider. The translation unit pulls the probe along the workpiece; the tip follows the valleys, while the glider slides over the peaks and serves as a reference; the electronics filter the signal and calculate the parameters.
  • Stylus: a small diamond cone with an angle of 60° or 90°, ending in a spherical tip with a nominal radius of 2 µm or 5 µm (10 µm for very coarse surfaces). Diamond does not wear down easily and does not scratch steel, because the measuring force is minimal: less than one millinewton for the 2 µm tip.
  • Transducer: converts the vertical displacement of the probe into an electrical signal. The most common type is inductive; laboratory instruments also use optical interference transducers. Vertical resolution reaches the nanometer or a few nanometers, with a range of a few hundred micrometers.
  • Shoe or external reference point: provides the line against which toe movements are measured (paragraph 36.4).
  • Translation unit: a motor with a precision guide that moves the probe at a constant speed, usually between a few tenths of a millimeter and 1 mm per second. An encoder indicates the probe tip’s location at all times.
  • Electronics: amplifies the signal, samples it at a constant rate, applies filters, calculates parameters, displays the profile, and generates the report. In portable devices, it is housed in a unit with a display; in benchtop instruments, it is a PC.

Angles, forces, speeds, and resolutions are approximate values: the manufacturer's specifications and ISO 3274, which describes stylus-type instruments, are authoritative.

The Cape and the Narrow Valleys

The tip has finite dimensions, so it does not fit everywhere: in a valley narrower than its diameter, it stops before reaching the bottom. The roughness tester always measures the profile “seen” by the tip, which is slightly rounded in the valleys. It is a mechanical filter, which is why the standard links the tip radius to the cutoff (table in paragraph 36.6). A chipped or worn tip, which has a radius larger than the nominal value, measures roughness values lower than the actual values.

tip with r = 2 µm6 µmreaches thebottom (3 µm)tip with r = 5 µm6 µm1 µm remainsbelow the edgetrace recorded by the instrument (lowest point of the tip)
Figure 36.3 · A groove 6 µm wide and 3 µm deep viewed with two probes. The 2-µm probe touches the bottom; the 5-µm probe hovers above the edges and detects a valley only 1 µm deep. Vertical and horizontal scales are the same.
The tip sinks into a narrow groove
h = r − √(r² − (w / 2)²) valid if w < 2 · r
h is the distance by which the tip extends below the edges of the groove, in µm; r is the tip radius in µm; w is the groove width in µm. If the groove is wider than the tip diameter (w ≥ 2r), the entire tip enters the groove and measures the cone angle.
Example 36.1 · How deep does the tip go?

A groove that is w = 6 µm wide and 3 µm deep. How much does a 5 µm tip sink? And a 2 µm tip?

5 µm tip: w / 2 = 3 µm < r → h = 5 − √(25 − 9) = 5 − √16 = 5 − 4 = 1 µm
2 µm tip: 2 · r = 4 µm < 6 µm → the entire cap fits into the groove
Inspection of the 60° cone at the edge height (Figure 36.3): The cone is approximately 5.8 µm wide, which is less than the 6 µm width of the groove, so the tip touches the bottom
5 µm tip: valley reading of 1 µm instead of 3 µm · 2 µm tip: full valley reading

36.4 Skate or external reference

To measure how much the tip rises and falls, you need a reference line. There are two options, each with different uses.

With a skatethe skate rests on the spikesrecorded profile: roughness onlyWith external referenceprecision straight-line guidanceshape, waviness, and roughness
Figure 36.4 · Above, the probe glides over the peaks and follows the long waves: the instrument detects only the roughness. Below, the probe moves along a straight guide and records the entire profile—including shape and undulation—which the filter then separates.

The Skate

The glide is a rounded support with a radius much larger than that of the tip, mounted next to the tip (in front, behind, or to the side). It glides over the peaks and follows the shape and long waves of the workpiece, so the tip measures only the deviations relative to the glide. The skid-type probe is compact, does not require careful leveling, and is used in portable roughness testers. However, it has some limitations:

  • It measures only roughness: it does not provide data on undulation, shape, steps, or the heights of a shoulder;
  • If the wave has a length close to the distance between the tip and the tail, the tail rises while the tip drops, and the profile becomes distorted;
  • If there is a hole or an isolated peak beneath the skid, a defect appears in the profile that does not actually exist;
  • On short curved surfaces (fillets, small spheres), the glider does not sit properly.

The external reference

In benchtop instruments—and in some high-end portable models—the probe is carried by a carriage that slides along a precision linear guide. The instrument records the complete primary profile and then, using filters, calculates roughness, waviness, and form. This is the most accurate method and allows for the measurement of steps, fillets, and profiles as well, but it requires careful leveling of the workpiece (section 36.8) and a stable workbench.

📌
Which one to choose. For Ra and Rz on flat or cylindrical surfaces of a certain size, the slide works very well and is faster. For surfaces with “max,” for Rt and Rmax on seals, for short or curved surfaces, and for any waviness measurement, the external reference is used.

36.5 The Filter: Profile, Waviness, and Roughness

The recorded profile contains everything: the shape of the workpiece, the long waves caused by vibrations and deflections, and the short tool marks (section 16.1). To calculate roughness, only the latter need be considered. The instrument does this using two filters:

  • The λs filter removes very short-period irregularities (noise, details finer than the tip can resolve) and yields the primary profile P;
  • The λc filter (the cutoff) calculates the average wavy line of the primary profile: this line is the waviness W; what remains, P − W, is the roughness profile R.
P-profile(primary)W-corrugation(average line)Roughness RR = P − Wλc: shorter waves go into R, longer ones into W
Figure 36.5 · At the top, the primary profile with its centerline (light blue dashed line). In the middle, the centerline alone—that is, the undulation. At the bottom, the difference: the roughness profile, which fluctuates around zero.

The Gaussian Filter

Today, the standard filter is the Gaussian filter (ISO 16610-21). For each point in the profile, it calculates a weighted average of the neighboring points, using bell-shaped weights: nearby points carry a lot of weight, while distant points carry almost none. The width of the bell curve is related to the cutoff. A wave much shorter than λc is “smoothed out” by the average and ends up entirely in the roughness; a wave much longer than λc passes into the mean line and ends up in the undulation. A wave exactly λc long is split in half.

Gaussian Filter Transfer Function
TW = 0,5(λc / λ)² TR = 1 − TW
TW a fraction of a wave's amplitude that results in a ripple · TR fraction that results in roughness · λ length of the irregularity in mm · λc Cut-off in mm. The standard writes the formula as e−π(α·λc/λ)² with α = √(ln 2 / π) = 0.4697: it's the same thing written another way.
0%50%100%0.080.250.82.58λc = 0.8 mm: half and halfpasses through the roughnesspasses through the wavinesslength of wave λ (mm), logarithmic scaletransmission
Figure 36.6 · The Gaussian filter with λc = 0.8 mm. Short waves (left) pass through the roughness, while long waves (right) pass through the waviness; the two curves intersect at 50% precisely at λc.
Example 36.2 · What the filter lets through

It is measured at λc = 0.8 mm. What fraction of three different waves is accounted for by the roughness?

turning grooves with a pitch of 0.2 mm: λc / λ = 4 → TW = 0.516 = 0.000015 → TR ≈ 100%
0.8 mm wavelength: λc / λ = 1 → TW = 0.5 → TR = 50%
2.4 mm wave length: λc / λ = 1/3 → TW = 0.51/9 = 0.926 → TR = 7.4%
The grooves play back perfectly, with almost no long-term ripple: the filter does what it's supposed to, as long as the cutoff is set correctly
📜
The old 2RC filter. Analog roughness meters used an electrical two-cell resistive-capacitive (2RC) filter. It did a poorer job of separating the waves and shifted the peaks; for this reason, it was replaced by the Gaussian filter. Many instruments still offer this option for comparing results with older measurements: the report must always specify which filter was used.

36.6 Cut-off and Measurement Lengths

Choosing the cutoff is the most important decision before taking a measurement. A cutoff that is too short cuts off part of the roughness and yields values that are too low; one that is too long allows waviness to be included in the roughness and yields values that are too high. The table in Section 16.2 listed cutoff values as a function of Ra; here we supplement it with Rz and the groove width Rsm, which is used for periodic profiles.

Cut-off λc = lr (mm)Expected Ra (µm)Expected Rz (µm)Rsm, periodic profiles (mm)ln (mm)
0,08more than 0.006 up to 0.02more than 0.025 up to 0.1more than 0.013 up to 0.040,4
0,25more than 0.02 up to 0.1more than 0.1 up to 0.5more than 0.04 up to 0.131,25
0,8more than 0.1 up to 2more than 0.5 up to 10more than 0.13 up to 0.44
2,5more than 2 and up to 10more than 10 up to 50more than 0.4 up to 1.312,5
8more than 10 up to 80more than 50 up to 200more than 1.3 up to 440

Guideline ranges taken from ISO 4288 (now incorporated into ISO 21920-3), to be used when the drawing does not specify a cutoff value. If the drawing specifies a cutoff value, the drawing takes precedence.

How to Choose

  • Periodic profile (turning, milling with regular pitch): Estimate Rsm, which for turning is nearly equal to the feed rate f, and select the cutoff value from the Rsm column. The bands are designed so that each base length contains at least two grooves.
  • Non-periodic profiling (grinding, lapping, EDM, sandblasting): Estimate Ra (from the drawing or by comparison with the reference plate), select the cut-off value, measure, and verify that the measured value falls within the range defined by that cut-off. If it falls outside the range, change the cutoff value and repeat the process.

The tip and sampling pitch must also be compatible with the cutoff frequency. ISO 3274 specifies the following combinations:

λc (mm)λs (µm)tip radius (µm)Maximum sampling step Δx (µm)
0,082,520,5
0,252,520,5
0,82,520,5
2,5851,5
825105

Approximate values according to ISO 3274. In practice, with λc = 0.8 mm, the 5 µm stylus is also often used, with a negligible error when Ra exceeds approximately 0.5 µm.

Base Length, Measurement Length, and Total Travel Length

The base length lr is equal to the cutoff; the evaluation length ln, unless otherwise specified in the drawing, is five base lengths (Section 16.2). However, the probe travels a longer distance: the Gaussian filter, in order to calculate the average line at the beginning and end, requires data points both before and after those points. Therefore, it travels a start-up stroke and an exit stroke, which are not included in the calculation.

Startoutputlr = λclr = λclr = λclr = λclr = λcλc/2λc/2evaluation length ln = 5 · lrtotal travel lt = ln + λc = 6 · λc
Figure 36.7 · The probe's travel with a Gaussian filter: half the start cutoff, five base lengths, half the end cutoff. The gray areas are used only for the filter.
Lengths, Points, and Measurement Time
ln = 5 · lr = 5 · λc lt = ln + 2 · λc2 = 6 · λc
N = lnΔx t = ltv
lr base length in mm · ln evaluation length in mm · lt total probe travel in mm · λc cutoff in mm · N number of points used for the calculation · Δx sampling step in mm · t travel duration in s · v translation speed in mm/s. The start and end of the half-cut-off each apply to the Gaussian filter; with the old 2RC filter, a full cut-off was required on each side (lt = 7 · λc). Some instruments use different travels: refer to the instrument manual.
Example 36.3 · Cut-off for a turned workpiece with a surface roughness of Ra 1.6

A pin is turned at f = 0.2 mm/rev and rε = 0.8 mm; the drawing specifies Ra 1.6 µm. What cutoff value should be used?

Theoretical roughness (Section 26.10): Ra ≈ 0.2² · 1000 / (31.2 · 0.8) = 40 / 24.96 = 1.60 µm
Periodic profile: Rsm ≈ f = 0.2 mm → range “greater than 0.13 up to 0.4” → λc = 0.8 mm
Check with Ra: 1.6 µm falls within the range “greater than 0.1 up to 2” → still λc = 0.8 mm
Each base length contains 0.8 / 0.2 = 4 grooves; ln = 5 × 0.8 = 4 mm
λc = 0.8 mm, ln = 4 mm, 2 µm tip (or 5 µm tip, since Ra exceeds 0.5 µm)
Example 36.4 · Cut-off for a ground workpiece and for roughing

A bore honed to a Ra of 0.2 µm and a hub rough-turned on a lathe with a feed rate of f = 0.5 mm/rev and a radius of 1.6 mm.

Grinding, non-periodic profile: Ra 0.2 µm in the range “greater than 0.1 up to 2” → λc = 0.8 mm, ln = 4 mm
Note: 0.2 µm is close to the edge of the band; if the measurement is less than 0.1 µm, repeat the measurement with λc = 0.25 mm
Roughing: Theoretical Ra = 0.5² · 1000 / (31.2 · 1.6) = 250 / 49.92 = 5.0 µm; Rsm ≈ 0.5 mm → range “greater than 0.4 up to 1.3” → λc = 2.5 mm
Inspection with Ra: 5.0 µm in the "over 2 to 10" range → λc = 2.5 mm, ln = 12.5 mm
Finished: λc 0.8 mm · Rough-machined: λc 2.5 mm (5 µm tip)
Example 36.5 · How long is the travel, and how long does it last?

With λc = 0.8 mm, a Gaussian filter, and a speed of 0.5 mm/s. And what about the old 2RC filter?

ln = 5 × 0.8 = 4.0 mm; lead-in and lead-out 0.8 / 2 = 0.4 mm each
lt = 4.0 + 0.4 + 0.4 = 4.8 mm (i.e., 6 × 0.8)
t = 4.8 / 0.5 = 9.6 s, plus the probe's return stroke
with the 2RC: lt = 7 × 0.8 = 5.6 mm
You need at least 4.8 mm of clear space and about 10 seconds per measurement
Example 36.6 · How many points does the instrument read?
λc = 0.8 mm: ln = 4 mm = 4000 µm, Δx = 0.5 µm → N = 4000 / 0.5 = 8000 points (9600 over the entire 4.8 mm travel)
λc = 2.5 mm: ln = 12.5 mm = 12,500 µm, Δx = 1.5 µm → N = 12,500 / 1.5 ≈ 8,333 points
About 8,000 points in both cases: the step size increases with the cutoff, while the number of points remains similar
Example 36.7 · A Shoulder that Is Too Short

A turned shoulder with a Ra of 1.6 µm is only 3 mm wide. Can it be measured with λc = 0.8 mm?

Normal travel: lt = 4.8 mm > 3 mm → that doesn't work
The number of base lengths is reduced: space for the bases = 3 − 0.8 (start and end) = 2.2 mm → 2.2 / 0.8 = 2.75 → 2 base lengths
ln = 2 · 0.8 = 1.6 mm; lt = 1.6 + 0.8 = 2.4 mm ≤ 3 mm
Measure using two baseline lengths and record the results in the report; the cutoff remains 0.8 mm, because it is based on the grooves and not on the space.
⚠️
Do not shorten the cutoff to make the measurement fit. On a short surface, it’s tempting to switch to λc = 0.25 mm: the travel becomes 1.5 mm, and everything fits. But if the grooves are spaced 0.2 mm apart, a base length of 0.25 mm contains just over one groove, and the filter cuts off part of it: the value ends up being too low. It’s better to have fewer base lengths with the correct cutoff.

36.7 Direction and Position of the Measure

Perpendicular to the furrows

Unless otherwise specified in the drawing, the measurement is taken perpendicular to the direction of the grooves (the direction of the grooves on the symbol is explained in paragraph 16.6). In this direction, the probe passes through all the grooves, and the measured roughness is the highest: this is the worst-case scenario, the one that matters for the function. When measuring along a groove, the tip slides along the bottom and records only the small irregularities of the cutting edge: the number is much lower and makes a surface appear good when it is not.

ABTurned workpiece viewed from above: the grooves are verticalA, perpendicular to the grooves: Ra 1.6 µmB, along a groove: Ra 0.3 µmsame surface, five times lower value:measurement B makes a workpiece appear good when it is not
Figure 36.8 · On the same turned workpiece, travel A—along the axis and thus perpendicular to the grooves—reveals all the grooves; travel B—along a groove—reveals only the bottom.

On a turned workpiece, the grooves run around the axis, so the correct travel direction is along the generatrix, parallel to the axis. On a faced surface, the grooves are circles (or a spiral), and the correct travel direction is radial. On a face-milled surface, the grooves are arcs: the measurement is taken in the direction of feed, at the center of the pass. Surfaces without a defined direction (sandblasting, erosion, cross-lapping) can be measured in any direction; as a precaution, measurements should be taken in two directions at 90°.

Where and how often

  • Measure where you expect the worst value: experience and the reference plate will help you find it (usually near the shoulders, where the tool slows down, or where the workpiece is more flexible and vibrates);
  • On a uniformly finished surface, start with one measurement and increase it in accordance with the acceptance rule (paragraph 36.10);
  • On shafts, measurements are taken along multiple generatrices—for example, three at 120° intervals—at the beginning, middle, and end of the section;
  • For small surfaces, the number of base lengths is reduced, and this is stated (example 36.7);
  • In holes, probes with long, thin arms are used; with special tips, they can enter holes just a few millimeters in diameter, but the depth they can reach is limited;
  • For radii and curved surfaces, an external reference and a filter that removes the shape are used; with the probe, measurements are taken only along a straight generatrix;
  • Keep at least a few tenths of a millimeter away from edges, chamfers, holes, and edges: if the probe touches an edge, it will give nonsensical readings.
beginning, middle, and end, along the axisshort segment123three generatricesat 120°hole: armlong and thin
Figure 36.9 · On the left, a shaft showing the travels along the axis (red arrows) at the beginning, middle, and end of the large diameter and along the short section. In the center, the three generatrices at 120°. On the right, the measurement in a hole with the thin arm.

36.8 Step-by-Step Implementation

A roughness measurement takes only a few seconds, but preparation is more important than the measurement itself. Here is the procedure to follow, whether using a portable or benchtop roughness tester.

  • Cleaning. Remove oil, emulsion, chips, and dust using a lint-free cloth and a suitable solvent. Never clean with sandpaper or a wire brush: this alters the surface you intend to measure. A speck of dust under the tip becomes a peak of a few micrometers.
  • Instrument Check. At the beginning of the shift (or before an important series of measurements), measure the supplied reference sample and compare the value with that on its certificate (paragraph 36.14). If the deviation exceeds the specified limit, adjust the instrument or send it for calibration.
  • Support. The workpiece must be held steady: on a flat surface, in a vise, or on a V-block for shafts. Keep it away from presses, compressors, and rough-machining tools: floor vibrations will be transmitted to the workpiece. The portable machine rests on its own feet or on a V-block; do not hold it in your hand while it is traveling.
  • Leveling and Positioning of the Probe. The travel must be parallel to the surface so that the probe tip remains within its measuring range. Using the external reference, level the workpiece (or table) until the position indicator remains centered throughout the entire travel; when using the slide, simply ensure that the probe is positioned straight and that the probe tip starts at the midpoint of the range.
  • Selection of parameters. Cut-off and number of base lengths (section 36.6), Gaussian filter, parameters to be calculated (those from the drawing, plus Rz and Rt, which cost nothing), acceptance rule.
  • Measurement. Start the travel without touching either the workpiece or the workbench.
  • Read and check the profile. Before entering the number, look at the profile: a very high isolated peak indicates dust or a burr; a profile that rises entirely on one side indicates improper leveling; a long, regular wave may indicate vibration. If the profile looks suspicious, clean it and repeat the process.
  • Repeat and record. Take measurements at the other designated points, note the values along with their locations, and save the report.

36.9 The measurement ratio and how to interpret it

The result of a measurement is a report (displayed on-screen, printed, or saved as a file) consisting of three parts: the measurement conditions, the profile graph, and the parameter values. Without the conditions, the numbers cannot be compared to anything.

Measurement Report · Shaft Ø40, surface A, generatrix 1λc 0.8 mm · ln 4 mm (5 × 0.8) · Gaussian filter · λs 2.5 µm · 2 µm 60° tip · v 0.5 mm/s ·0.5 mm2 µmRa1.22 µmRq1.46 µmRz5.8 µmRmax6.3 µmRt6.7 µmRsm0.150 mmRmr (1.0 µm)10%Drawing: Ra 1.6 µm (16% rule) · turning f 0.15 mm/rev, rε 0.8 mmMeasurement result: 1.22 ≤ 1.6, value below the limit
Figure 36.10 · A typical measurement report. At the top are the conditions; on the left is the roughness profile (the vertical scale is much larger than the horizontal scale); on the right are the parameters. The regular grooves from turning are clearly visible.
  • Conditions: cut-off values for λc and λs, number of base lengths and evaluation length, filter, nose radius and angle, slide or external reference, speed, date, tool, operator, position on the workpiece.
  • Profile: Keep in mind that the two scales are different. In the figure, the vertical magnification is about 90 times that of the horizontal magnification (compare the two scale bars, 2 µm and 0.5 mm). The grooves look like sharp saw teeth, but in reality they are very wide and very shallow depressions.
  • Ra, Rq, Rz, Rt: defined in Chapter 16 (Sections 16.3 and 16.4).
  • Rmax (or Rz1max): the highest of the five Zt heights, useful for sealing surfaces.
  • Rsm: the average width of the profile elements, that is, the average distance between one peak and the next, in mm. On a turned workpiece, it is almost equal to the feed rate: in the figure, 0.150 mm with f = 0.15 mm/rev. If Rsm differs from the feed rate, something is wrong (vibrations, cutting edge).
  • Rmr(c): the lift-to-drag ratio, explained below.

How the tool calculates the parameters

To understand how the electronics work, let’s redo the calculation by hand for a few points. The instrument uses thousands of them, but the procedure is the same: find the average line, measure the ordinates from that line, and apply the definitions from Chapter 16.

−2−10123z (µm)average linez = +0.8 µmreferencez = 0Ordinates from the mean line (µm):1.12.4−0.1−1.60.31.7−1.1−2.8−0.41.42.60−1.9−0.70.7−1.6
Figure 36.11 · The sixteen data points from Example 36.8. The readings start at any zero point (dashed line); the mean line (blue) is 0.8 µm higher. Below, the ordinates measured from the mean line.
Example 36.8 · Ra from a 16-point sampled profile

The transducer provides sixteen readings at constant intervals, measured from any zero point (in µm): 1.9 · 3.2 · 0.7 · −0.8 · 1.1 · 2.5 · −0.3 · −2.0 · 0.4 · 2.2 · 3.4 · 0.8 · −1.1 · 0.1 · 1.5 · −0.8. Calculate Ra, Rp, Rv, and Rt for this section.

Sum of the readings: 1.9 + 3.2 + 0.7 − 0.8 + 1.1 + 2.5 − 0.3 − 2.0 + 0.4 + 2.2 + 3.4 + 0.8 − 1.1 + 0.1 + 1.5 − 0.8 = 12.8
average line: 12.8 / 16 = 0.8 µm → subtract 0.8 from each reading
y-coordinates relative to the mean: 1.1 · 2.4 · −0.1 · −1.6 · 0.3 · 1.7 · −1.1 · −2.8 · −0.4 · 1.4 · 2.6 · 0 · −1.9 · −0.7 · 0.7 · −1.6
Check: Their sum is zero (positive 10.2, negative −10.2) ✓
sum of the absolute values: 10.2 + 10.2 = 20.4 → Ra = 20.4 / 16 = 1.275 µm
Highest peak Rp = 2.6 µm; deepest valley Rv = 2.8 µm → Rt = 2.6 + 2.8 = 5.4 µm
Ra ≈ 1.28 µm · Rt = 5.4 µm (Rt / Ra = 4.2, a typical value for a machined surface)
Example 36.9 · Rz, Rmax, and Rt from Five Base Lengths

Across the five base lengths, the roughness meter detects these highest peaks (Zp) and deepest valleys (Zv) (in µm, from the mean line):

Section12345
Zp2,63,12,43,52,8
Zv2,92,53,32,73,6
Zt = Zp + Zv for each segment: 5.5 · 5.6 · 5.7 · 6.2 · 6.4 µm
Rz = (5.5 + 5.6 + 5.7 + 6.2 + 6.4) / 5 = 29.4 / 5 = 5.88 µm
Rmax = the largest Zt = 6.4 µm (segment 5)
Rt = highest peak of the entire measurement + deepest trough = 3.5 (section 4) + 3.6 (section 5) = 7.1 µm
Rz = 5.88 µm · Rmax = 6.4 µm · Rt = 7.1 µm
Example 36.10 · Theoretical roughness and measured roughness

The ratio shown in Figure 36.10 is based on a turned shaft with f = 0.15 mm/rev and rε = 0.8 mm. Compare the measured values with the theoretical values from the formula in Section 26.10.

Theoretical Ra = 0.15² · 1000 / (31.2 · 0.8) = 22.5 / 24.96 = 0.90 µm
Theoretical Rt = 0.15² · 1000 / (8 · 0.8) = 22.5 / 6.4 = 3.52 µm
Measured values: Ra = 1.22 µm → 1.22 / 0.90 = 1.35 times the theoretical value; Rz = 5.8 µm → 5.8 / 3.52 = 1.65 times
Measured RSM: 0.150 mm = f: The grooves are those caused by feed; there are no noticeable vibrations.
The actual surface is about one-third rougher than the theoretical surface: this is normal (wear, material deformation, minor vibrations). Rz increases more because it detects individual scratches
🔧
Use the ratio to adjust the machine. The ratio between the measured and theoretical roughness for a given machine, with a specific material and a specific insert, remains fairly constant. Once measured (here, 1.35), it can be used to select the feed rate: to achieve a measured Ra of 1.6 with a margin, aim for a theoretical Ra of approximately 1.6 / 1.35 = 1.19 µm, and then calculate f using the inverse formula in Section 26.10.

The Abbott Curve and the Lift-to-Drag Ratio

Let’s imagine cutting through the profile with a horizontal line at a depth c below the highest peak. The line passes through sections of material and sections of void. The sum of the sections of material, divided by the evaluation length, is the lift ratio Rmr(c): It indicates how much surface area a resting plane would touch after the peaks have been worn down by c. By varying c from the highest peak to the deepest valley, we obtain the Abbott curve (or lift curve), which ranges from 0% to 100%.

Lift-to-drag ratio
Rmr(c) = Ml(c)ln × 100
Rmr(c) lift ratio in % · c depth of the cutting level, measured from the highest peak (or another specified reference level), in µm · Ml(c) sum of the lengths of the material segments cut by the level, in mm · ln evaluation length in mm. For a sampled profile, Ml/ln is the fraction of points that lie above the level.
cprofile and cutting level0%50%100%Rmr(c) = 38%Abbott curvelift ratio Rmr
Figure 36.12 · On the left, a profile with smoothed peaks (like a polished cylinder liner) cut at level c: the sections of material are shown in red. On the right, the Abbott curve for the same profile: level c corresponds to the indicated lift-to-drag ratio.
Example 36.11 · Lift-to-drag ratio from the sixteen points

Using the ordinates from Example 36.8 (highest peak 2.6 µm), calculate Rmr for c = 1 µm and c = 3 µm.

sorted in descending order: 2.6 · 2.4 · 1.7 · 1.4 · 1.1 · 0.7 · 0.3 · 0 · −0.1 · −0.4 · −0.7 · −1.1 · −1.6 · −1.6 · −1.9 · −2.8
c = 1 µm → level 2.6 − 1 = 1.6 µm: the levels above are 2.6, 2.4, and 1.7 → 3 points out of 16 → RMR = 3 / 16 × 100 = 18.75%
c = 3 µm → level 2.6 − 3 = −0.4 µm: there are 9 points above (or equal to) this level → Rmr = 9 / 16 × 100 = 56.25%
Rmr(1 µm) ≈ 19% · Rmr(3 µm) ≈ 56% (with thousands of data points, the instrument provides more precise values, but the reasoning is as follows)

Two surfaces with the same Ra value can have very different Abbott curves (Figure 16.6): a surface with plateaus and narrow valleys has a high bearing ratio even just a few micrometers from the peak, carries the load well, and retains oil in the valleys. For these surfaces, such as cylinder liners and guides, the standard also defines the parameters Rpk (height of the peaks that will wear down during break-in), Rk (depth of the load-bearing core), and Rvk (depth of the valleys that retain the lubricant), derived from the Abbott curve. Here, it is sufficient to know that these parameters exist and that they can be found in the report when specified in the drawing.

36.10 Comparison with the drawing: the 16% rule and the maximum rule

A surface does not have “a single” roughness value: measurements taken at different points yield different values. To determine whether the workpiece is in compliance, a rule is needed, which depends on how the value is specified in the drawing (section 16.6).

  • 16% Rule (the default rule in ISO 1302 and ISO 4288, when only “Ra 1.6” is specified): The surface is acceptable if no more than 16% of the measured values exceed the limit.
  • Maximum Rule (when it says “Ra max 1.6”): No measured value, across the entire surface, must exceed the limit. Measurements must be taken at least three times at the points where the highest values are expected.

With the ISO 21920 series, the acceptance criteria have been revised: for new drawings, it is advisable to check the standard referenced in the title block and, if in doubt, ask the engineering department which criterion applies. Here we describe the two standard criteria, which are the ones found on the vast majority of drawings.

The 16% Rule Procedure

ISO 4288 provides a step-by-step procedure that omits certain measurements when the surface is clearly in good condition. In simplified form:

  • Take an initial measurement at the point that appears to be the worst: if the value does not exceed 70% of the limit, the surface is accepted;
  • Otherwise, two more measurements are taken: if none of the three values exceeds the limit, the surface is accepted; if two or more values exceed the limit, it is rejected;
  • If even one value exceeds the limit, the total number of measurements reaches twelve: the area is accepted if no more than two of the twelve values exceed the limit.

Educational summary prepared by CNC Simulator Italia; for contractual acceptance testing, the text of the standard referenced on the drawing shall prevail.

Example 36.12 · How Many Measurements Are Needed?

A shaft has a designed Ra of 0.8 µm, with the 16% rule. 70% of the limit is 0.7 × 0.8 = 0.56 µm. Three different workpieces.

Workpiece A: first measurement 0.49 µm ≤ 0.56 → accepted based on a single measurement
Workpiece B: first measurement 0.63 µm (79% of the limit) → two others: 0.58 and 0.74 µm, all ≤ 0.8 → accepted based on three measurements
Workpiece C: 0.63 · 0.85 · 0.77 µm → one value exceeds 0.8 → this results in twelve: 0.69 · 0.72 × 0.66 × 0.71 × 0.75 × 0.62 × 0.78 × 0.70 × 0.68 → of the twelve, only 0.85 exceeds the threshold
One value out of twelve is incorrect: 1 ≤ 2 → accepted
A: 1 measure · B: 3 measures · C: 12 measures—all accepted
1.01.21.41.61.8123456789101112limit 1.6mean + saverageMeasurement number (Ra in µm)
Figure 36.13 · The twelve measurements from Example 36.13 (vertical axis in 1.0 µm increments). Two bars exceed the 1.6 µm limit. The green line—the mean plus one standard deviation—remains below the limit.
Example 36.13 · The 16% Rule over Twelve Measures

On a surface with a surface roughness (Ra) of 1.6 µm, twelve values (in µm) are measured: 1.32 · 1.45 · 1.18 · 1.52 · 1.71 · 1.39 · 1.27 · 1.48 · 1.66 · 1.35 · 1.41 · 1.29. Does the surface meet the specifications?

values above 1.6: 1.71 and 1.66 → 2 out of 12, or 16.7%
The procedure allows for up to 2 values out of 12 (2 out of 12 is 16.7%, the whole number closest to 16% of 12, which is 1.92) → compliant
To understand where the 16% comes from: mean = 17.03 / 12 = 1.419 µm; standard deviation s = 0.156 µm (formula from Section 31.6)
average + s = 1.419 + 0.156 = 1.575 µm ≤ 1.6
Correct. In a normal distribution, 15.9% of the values lie above the mean plus one standard deviation: the 16% rule essentially requires that the mean plus s not exceed the limit.
Example 36.14 · The Maximum Rule and Uncertainty

The same twelve measurements, but the drawing says “Ra max 1.6.” Then there is another workpiece marked “Ra max 0.8,” measured with an expanded uncertainty U = 0.06 µm, which gives a maximum of 0.71 µm; a third workpiece gives a maximum of 0.77 µm.

Maximum rule: 1.71 and 1.66 exceed 1.6 → non-compliant, even though the other ten values are acceptable
Second workpiece: 0.71 + 0.06 = 0.77 µm ≤ 0.8 → compliance proven even when taking uncertainty into account (paragraph 31.9)
Third workpiece: 0.77 ≤ 0.8, but 0.77 + 0.06 = 0.83 µm > 0.8 → the value falls within the uncertainty range
Same numbers, different rule, opposite result. Given the uncertainty, below 0.8 − 0.06 = 0.74 µm, compliance is certain; between 0.74 and 0.86, there is still uncertainty.
📐
How large is the uncertainty of a roughness measurement? Proportionally, much greater than for a diameter. Between instrument calibration, repeatability, and—above all—variation from one point to another on the surface, an uncertainty of 5–10% of the value is normal in a workshop. The “one-tenth” rule in paragraph 31.8 does not apply strictly here: the uncertainty is compared to the limit, and a margin is maintained.

36.11 Common Errors

Almost all errors in roughness measurements have the same effect: a number that appears precise—with three digits—but does not accurately describe the surface. These are the most common ones.

ErrorWhat's happening?Remedy
Measurement parallel to the furrowsvalues that are much lower than the actual values (Figure 36.8)perpendicular to the grooves, unless otherwise specified in the drawing
Worn or chipped tipThe actual radius increases, the tip does not enter the valleys, and Ra and Rz decrease; chipping may also cause scratches on the workpiecePeriodic inspection of the tip using the sample or under a microscope; replacement
VibrationsRegular waves or noise in the profile, high valuesDamped bench, installed away from heavy machinery, no contact during travels
Skating on a curved or short surfaceThe skate rises and falls with the curve; the profile deforms.external reference with a shape filter, or travels along a straight generatrix
Cut-off is too shortPart of the roughness is accounted for by waviness: low valuesSelecting the cutoff value using the table in paragraph 36.6 and verifying the range
Cut-off is too longWaviness contributes to roughness: high valuesas above
Undeclared reduced evaluation lengthless stable values that cannot be compared with other measurementsAlways include the number of base lengths on the report
Dirty or oily workpieceIsolated peaks (powder) or filled valleys (oil): distorted Rz and Rt valuesThorough cleaning; check the profile before reading the numbers
Travel that touches an edge or a holehuge spikes in the graph, nonsensical valuesto start and end far from the edges
Off-screen tip due to incorrect levelingprofile cut at the top or bottom, error message, or low valuesLevel the workpiece; center the drill bit within the area
A filter different from the reference filter (2RC instead of Gaussian)differences of a few percent—sometimes more—compared to the customer's measurementsUse the Gaussian filter and include it in the report
Example 36.15 · The 0.8 Cut-off Rule

The hub from Example 36.4 (theoretical Ra of 5.0 µm, grooves every 0.5 mm) is routinely measured with λc = 0.8 mm instead of 2.5 mm. What happens to the main component of the profile—the one corresponding to the groove pitch?

λc / λ = 0.8 / 0.5 = 1.6 → (λc / λ)² = 2.56
TW = 0,52,56 = 0,170 → TR = 1 − 0,170 = 0,830
Nearly one-sixth of the groove width is accounted for by the waviness; if the profile consisted solely of that wave, Ra would change from 5.0 to approximately 5.0 × 0.83 = 4.2 µm
with λc = 2.5 mm: λc / λ = 5 → TW = 0.525 ≈ 0 → all the grooves pass through
With the wrong cut-off, the workpiece appears about 17% smoother: enough to deem the surface a reject

36.12 Portable, benchtop, and in-process roughness meters

Laptops

The portable roughness tester features a translation unit and a sliding probe in a small, battery-powered housing that can often be detached from the display. It is taken to the machine, placed on the workpiece—even while it is still mounted—measures Ra, Rz, Rt, Rmax, Rsm, and a few other parameters, and prints or saves the results. It is the workshop’s go-to tool: fast, rugged, and with a measurement uncertainty suitable for most inspections. With magnetic mounts, V-blocks, and extensions, it adapts to holes, grooves, and vertical surfaces.

Countertop

The benchtop roughness tester (or profilometer) features a motorized column, a table that can often be tilted for leveling, long travels (up to tens or hundreds of millimeters), and an external reference. It measures the complete primary profile—including waviness, shape, radii, and angles—and calculates all parameters, including the Abbott curve and Rk parameters. It is housed in the metrology lab on a damped bench and serves as the instrument for testing and dispute resolution.

In the car and in production

  • Roughness probes on coordinate measuring machines (Chapter 37): The machine positions the probe at the correct location and orients it perpendicular to the grooves, even inside holes; dimensions and roughness are measured with a single setup.
  • Roughness probes in machine tools: Mounted in the spindle like a measuring probe, they inspect the freshly machined surface before the workpiece is removed. They require a stationary, clean machine and a dedicated measurement cycle.
  • Optical light-scattering sensors: A beam strikes the surface, and a sensor measures how the reflected light scatters. In a fraction of a second, they provide a roughness index without making contact; they are used to inspect 100% of workpieces on the production line, after comparing the index with measurements taken using a stylus.
  • Statistical control: The values measured during production are recorded on control charts (Chapter 38). A slowly increasing Ra value indicates tool wear; a sudden jump indicates chipping or a change in material.

36.13 3D Measurements and Noncontact Methods

From the profile to the surface

A profile is a line: it describes the surface only along the path of the probe tip. A ground, sandblasted, or eroded surface may have isolated defects (a crater, a scratch) that one travel detects but another does not. Area measurements, on the other hand, cover an entire area—for example, a square a few millimeters in size—much like a map of heights. The parameters are defined by ISO 25178 and are identified by the letter S:

  • Sa: arithmetic mean of the absolute heights over the area, the areal “counterpart” of Ra;
  • Sq: standard deviation of the heights (same as Rq);
  • Sz: maximum height of the surface, from the highest peak to the deepest valley in the entire area;
  • Other parameters describe the direction of the grooves, the actual surface area, and the volumes of material and void space, which are important for sealing and lubrication.

Sa and Ra are not the same thing: on surfaces without a specific direction, they are similar; on surfaces with oriented grooves, Ra depends on the direction of travel, while Sa does not. Different parameters are specified on the drawing, and they cannot be interchanged without the designer’s approval.

Arithmetic Mean Area Height
Sa = 1A ∬ |z(x, y)| dx dy Sa ≈ |z1| + |z2| + ... + |zn|n
Sa in µm · A measured area · z(x, y) height of a point above the mean plane, in µm · n number of grid points
Example 36.16 · Sa and Sz from a 16-point grid

A 4 × 4 grid of heights (µm), measured from any plane: first row 0.6 · −0.4 · 1.2 · 0.2; second row −0.9 · 0.3 · −0.2 · 1.0; third row 0.5 · −1.3 · 0.8 · −0.5; fourth row 0.1 · 0.9 · −0.7 · 0.4.

sum = 2.0 → average value at 2.0 / 16 = 0.125 µm, to be subtracted from all values
heights above the mean plane: 0.475 · −0.525 · 1.075 · 0.075 · −1.025 · 0.175 · −0.325 · 0.875 · 0.375 · −1.425 · 0.675 · −0.625 · −0.025 · 0.775 · −0.825 · 0.275
sum of absolute values = 9.55 → Sa = 9.55 / 16 = 0.597 µm
Sz = 1.075 − (−1.425) = 2.5 µm
Sa ≈ 0.60 µm · Sz = 2.5 µm (an optical instrument uses millions of points, but this is the basic calculation)

Non-contact methods

For area measurements, for very fine or soft surfaces (plastics, coatings, lenses), and for microstructures, non-contact optical instruments are used. The three most common principles are as follows.

Chromatic confocalwhite lightlensEach color comes into focus at a certain dimension:the returning color indicates the heightWhite-light interferometrycamerareferencereferencesplitterscanningthe fringes have maximum contrastwhen the spot is at the correct dimensionFocus variationcameralensImages are taken at different dimensions:for each point, the sharpest one is selected
Figure 36.14 · Three optical principles, greatly simplified. The vertical scanning motion is shown in red.
  • Confocal microscope: A small hole (pinhole) in front of the sensor allows only light from the points in focus to pass through. By moving the objective vertically, the position of maximum brightness is determined for each point, which corresponds to its height. In the color version, white light is used along with a lens that focuses each color at a different dimension: the color that is reflected back indicates the height, without moving anything.
  • White-light interferometry: the light splits into two beams, one of which strikes a workpiece and the other a reference mirror; when they recombine, they form light and dark fringes. With white light, the fringes are sharp only when the two paths are equal: during vertical scanning, the point of maximum contrast provides the dimensions of each pixel. The vertical resolution is less than one nanometer.
  • Focus-variation microscope: It takes many images at different dimensions and, for each pixel, searches for the image in which it is sharpest. It works well on rough and sloped surfaces, but it needs some roughness; on a mirror-smooth surface, it can’t find anything to focus on.
MethodStrengthsLimitations
Stilocomplies with standards and is comparable, sturdy, affordable, works well in the workshoponly profiles (or slow-scanning areas); a tip that filters and can scratch soft materials
ConfocalGlossy and matte surfaces, steep slopes, high lateral resolutionsmall field of view, cost, susceptible to reflections
White-Light InterferometryExtremely high vertical resolution, fast on fine surfacesDifficult to use on very rough or very steep surfaces; sensitive to vibrations
Focus Shiftrough surfaces, shapes and edges, toolsNot suitable for mirror-like surfaces; lower vertical resolution

Indicative comparison prepared by CNC Simulator Italia. Optical instruments and styluses do not always yield the same reading on the same surface: the probe and the optical system “see” differently. For inspection purposes, use the method specified in the drawing or agreed upon with the customer. Other modern measuring instruments are covered in Chapter 37.

36.14 Calibration of the roughness tester

Like any measuring instrument (paragraph 31.14), the roughness tester must be calibrated at regular intervals and checked frequently. Roughness calibration standards, as described in ISO 5436-1, are used and classified into types based on what they are designed to measure.

Type A · DepthCertified groove depthdType C · Pitch and AmplitudeRegular waves: Ra and Rsm knownpType D · roughnessirregular, repeating profilelrlr
Figure 36.15 · Three types of calibration samples (exaggerated profiles). Type A checks the vertical amplification; Type C also checks the horizontal scale and the calculation of parameters; Type D checks the instrument as a whole using a profile similar to the actual one.
  • Type A: Wide grooves with a flat (or rounded) bottom and a certified depth. They are used to calibrate the vertical amplification: the instrument must read the depth d.
  • Type B: Narrow grooves or thin edges, to check the condition and radius of the tip.
  • Type C: periodic waveforms (sine waves, triangles, arcs) with known period and amplitude; they also control the horizontal scale, the filter, and the calculation of Ra and Rsm.
  • Type D: irregular profiles, similar to those produced by grinding, that repeat every base length; they certify Ra and Rz and verify the instrument under conditions that most closely resemble actual use.
  • Type E: Shape gauges (arcs, prisms) for instruments that also measure profiles and radii.

True calibration is performed in an accredited laboratory, which issues a certificate listing the measured values and their uncertainty. In the workshop, every time the instrument is turned on or at the start of a shift, a check is performed using the calibration block supplied with the instrument (usually a block with a Ra of a few micrometers and its certified value): if the reading deviates too much, adjust the instrument’s gain according to its manual, or send it in for calibration. The standard must be kept clean, in its case, and always measured at the same point and in the same direction: the tip, due to repeated use, will eventually wear it down.

Example 36.17 · Verification Using the Reference Sample

The supplied sample has a certified Ra value of 2.97 µm. The company allows for a tolerance of ±5% (internal standard). Five readings taken this morning: 3.04 · 3.01 · 3.07 · 2.99 · 3.05 µm.

average = (3.04 + 3.01 + 3.07 + 2.99 + 3.05) / 5 = 15.16 / 5 = 3.032 µm
error = 3.032 − 2.97 = +0.062 µm → +0.062 / 2.97 × 100 = +2.1%
+2.1% is within ±5% → this is how you use the tool
If one were to correct this, the gain adjustment factor would be 2.97 / 3.032 = 0.980
Verified instrument: reads 2.1% higher, within the internal tolerance limit of 5%
🏷️
Verification is not the same as calibration. The morning check using a sample determines whether the instrument is still accurate for a single value; laboratory calibration checks it against multiple values, using traceable standards and a stated uncertainty. The two processes complement each other: daily verification detects faults between calibrations, such as a chipped tip.

36.15 The Formulas in This Chapter

WhatFormula or ruleUnits
Ra scores in league playRa = (|z1| + ... + |zn|) / n, where z is the meanµm
Rz, Rmax, Rt (Chapter 16)Rz = average of the 5 Zt values · Rmax = max(Zt) · Rt = Zp,max + Zv,maxµm
Sinking of the tiph = r − √(r² − (w/2)²)µm
Gaussian Filter Transfer FunctionTW = 0.5(λc/λ)² · TR = 1 − TWpure number
Evaluation Lengthln = 5 · lr = 5 · λcmm
Total Travel (Gaussian Filter)lt = ln + λc = 6 · λcmm
Number of points and timeN = ln / Δx · t = lt / vpoints · s
Lift-to-drag ratioRmr(c) = Ml(c) / ln × 100%
Theoretical Roughness in Turning (Chapter 26)Ra ≈ 1000 · f² / (31.2 · rε) · Rt = 1000 · f² / (8 · rε)µm
The 16% Rule (statistical interpretation)no more than 2 of the 12 values above the limit · approximately mean + s ≤ limitµm
Compliance with Uncertaintyvalue + U ≤ upper limitµm
Area parameter SaSa = (|z1| + ... + |zn|) / n, where z is from the mean planeµm
Validation Errore = (average − certified value) / certified value × 100%

36.16 Exercises

Try to solve them on your own, using a calculator, before looking at the solution.

Exercise 1. Twelve readings of a profile, starting from any zero value (µm): 0.9 · −0.4 · 1.7 · 0.3 · −1.2 · 0.6 · 2.1 · −0.9 · 0.2 · 1.4 · −1.5 · 0.4. Calculate the mean line, Ra, and Rt.
Solution

Sum 3.6 → average line 3.6 / 12 = 0.3 µm. Y-coordinates: 0.6 · −0.7 · 1.4 · 0 · −1.5 · 0.3 · 1.8 · −1.2 · −0.1 · 1.1 · −1.8 · 0.1. Sum of absolute values 10.6 → Ra = 10.6 / 12 = 0.88 µm. Rt = 1.8 + 1.8 = 3.6 µm.

Exercise 2. For the five base lengths, the highest peaks are 1.8 · 2.2 · 1.6 · 2.5 · 2.0 µm, and the deepest valleys are 2.1 · 1.7 · 2.4 · 1.9 × 2.6 µm. Calculate Rz, Rmax, and Rt.
Solution

Zt = 3.9 × 3.9 × 4.0 × 4.4 × 4.6 µm. Rz = 20.8 / 5 = 4.16 µm. Rmax = 4.6 µm. Rt = 2.5 + 2.6 = 5.1 µm.

Exercise 3. What cut-off value would you choose for: a lapped surface with an expected Ra of 0.04 µm; a ground surface with an expected Ra of 0.1 µm; a turning surface with f = 0.3 mm/rev and rε = 1.6 mm; a non-periodic peripheral milling operation with an expected Ra of 3.2 µm?
Solution

Lapping: range “greater than 0.02 up to 0.1” → 0.25 mm. Grinding (Ra 0.1): still within the range “greater than 0.02 up to 0.1” → 0.25 mm (pay attention to the edge). Turning: Theoretical Ra = 0.3² · 1000 / (31.2 · 1.6) = 90 / 49.92 = 1.80 µm; Rsm ≈ 0.3 mm in the range “over 0.13 up to 0.4” → 0.8 mm, confirmed by Ra 1.8 µm in the range “over 0.1 up to 2.” Non-periodic milling Ra 3.2: 2.5 mm.

Exercise 4. With λc = 2.5 mm and a Gaussian filter, how long is the probe’s travel distance? How long does it take at 1 mm/s? And with the 2RC filter?
Solution

ln = 5 × 2.5 = 12.5 mm; lt = 6 × 2.5 = 15 mm; t = 15 / 1 = 15 s. With the 2RC: lt = 7 × 2.5 = 17.5 mm, 17.5 s.

Exercise 5. Given λc = 0.25 mm and a sampling step of 0.5 µm, how many points are included in the calculation for five base lengths?
Solution

ln = 5 × 0.25 = 1.25 mm = 1250 µm; N = 1250 / 0.5 = 2500 points.

Exercise 6. If λc = 0.8 mm, what percentage of a 1.6 mm-long wave ends up in the roughness profile? And what about a 0.4 mm-long wave?
Solution

1.6 mm wavelength: λc / λ = 0.5 → TW = 0.5 × 0.25 = 0.841 → TR = 15.9%. 0.4 mm wavelength: λc / λ = 2 → TW = 0.54 = 0.0625 → TR = 93.75%.

Exercise 7. A groove is 3 µm wide. How far below the edges does a 2 µm tip extend? And a 5 µm tip?
Solution

w / 2 = 1.5 µm. 2 µm tip: h = 2 − √(4 − 2.25) = 2 − 1.323 = 0.68 µm. 5 µm tip: h = 5 − √(25 − 2.25) = 5 − 4.770 = 0.23 µm. Both remain suspended: no tip can measure such a narrow groove in its entirety.

Exercise 8. A surface with a Ra of 3.2 µm (16% rule) yields twelve measurements (µm): 2.71 · 3.05 · 2.88 · 3.34 · 2.62 · 3.41 · 2.95 · 3.12 · 2.79 · 3.27 · 2.84 · 3.01. Does it meet the requirement? Also check using the mean ± s.
Solution

Above 3.2, there are 3.34, 3.41, and 3.27: three values out of twelve, more than the two allowed → non-conforming. Mean = 2.999 µm, s = 0.250 µm, mean + s = 3.249 µm > 3.2: statistical analysis also indicates that the 16% limit has been exceeded.

Exercise 9. The drawing specifies “Ra max 0.4.” Three measurements at the worst points yield 0.36, 0.38, and 0.35 µm, with an expanded uncertainty U = 0.03 µm. Does the workpiece meet the specification? Below what value would compliance be proven?
Solution

All values are below 0.4, but the highest one—0.38 + 0.03 = 0.41 µm—exceeds the limit: compliance is not proven, and the value falls within the zone of doubt. Compliance is proven if all values are below 0.4 − 0.03 = 0.37 µm.

Exercise 10. A lathe finishing operation with f = 0.12 mm/rev and rε = 0.4 mm yields a measured Ra of 1.45 µm. Calculate the theoretical Ra and the measured-to-theoretical ratio. Given the same ratio, rounded to 1.26, what feed rate is required to achieve a measured Ra of 0.8 µm?
Solution

Theoretical Ra = 0.12² · 1000 / (31.2 · 0.4) = 14.4 / 12.48 = 1.15 µm; ratio 1.45 / 1.15 = 1.26. Required theoretical Ra = 0.8 / 1.26 = 0.635 µm; f = √(31.2 × 0.4 × 0.635 / 1000) = √0.00792 = 0.089 mm/revolution, or approximately 0.08 mm/revolution.

Exercise 11. The reference sample has a certified Ra value of 0.45 µm. Five readings: 0.41, 0.43, 0.42, 0.44, 0.41 µm. With an internal limit of ±5%, can the instrument be used?
Solution

Average = 2.11 / 5 = 0.422 µm; error = 0.422 − 0.45 = −0.028 µm = −6.2%. Outside the ±5% range: the instrument is reading too low. Check the tip (a worn tip produces low readings), clean the sample, and if the error persists, adjust the instrument or send it in for calibration.

Exercise 12. Ten ordinates of a profile relative to the centerline (µm): 1.2 · −0.3 · 0.8 · −1.4 · 0.2 · 1.6 · −0.7 · 0.5 · −1.1 · −0.8. Calculate Rmr for c = 1 µm and c = 2 µm below the highest peak.
Solution

Highest peak: 1.6 µm. c = 1 µm → level 0.6 µm: above this are 1.6, 1.2, and 0.8 → 3 out of 10 → RMR = 30%. c = 2 µm → level −0.4 µm: above this are 1.6, 1.2, 0.8, 0.5, 0.2, and −0.3 → 6 out of 10 → Rmr = 60%.

Exercise 13. A turned shoulder (grooves every 0.2 mm, expected Ra of 1.6 µm) is 2 mm wide. Given λc = 0.8 mm and a Gaussian filter, how many base lengths fit within it? Why is it not advisable to switch to λc = 0.25 mm?
Solution

Space for the bases: 2 − 0.8 = 1.2 mm → 1.2 / 0.8 = 1.5 → a single base length (ln = 0.8 mm, lt = 1.6 mm). With λc = 0.25 mm, the 0.2 mm grooves would be outside the Rsm range for that cutoff (up to 0.13 mm): each base would contain only 0.25 / 0.2 = 1.25 grooves, and the filter would cut out part of the roughness: the value would be too low. Measure using one base, record the result in the report, and repeat the measurement at multiple points along the shoulder.