Roughness Measurement
In Chapter 16, we learned how to read and write roughness on drawings: Ra, Rz, the classes, and the symbols. Here, we’re switching sides of the bench: how to actually measure it. We’ll see when it’s enough to run your fingernail over a reference sample and when you need a roughness tester, how the instrument is constructed, what the filter does, how to choose the cutoff, where and in which direction to place the probe, how to read the measurement ratio, and how to determine if the workpiece is acceptable. The chapter concludes with a discussion of the most common errors, optical instruments, and calibration.
36.1 Why Is Roughness Measured?
Roughness is not merely an aesthetic consideration. It determines whether a seal ring leaks oil, whether a guide wears out in a month or in ten years, whether a shaft breaks due to fatigue starting from the bottom of a groove, whether a interference fit holds, and whether paint adheres. The recommended roughness values for various functions are listed in the table in Section 16.10; here, we are concerned with the next step: verifying that the workpiece coming off the machine meets these specifications.
Measuring surface roughness is useful not only for inspectors but also for those who actually do the work. An Ra value that increases from workpiece to workpiece is the first sign of a worn insert, a worn cutting edge, or a spindle that is starting to vibrate. Those who monitor roughness can spot problems before they result in scrap.
When a comparison Is Enough, and When a Tool Is Needed
Not all surfaces require a roughness tester. For roughing, for a surface that doesn’t come into contact with anything, or for a quick check right at the machine, it’s enough to compare the workpiece with a sample—by sight and touch. The instrument becomes essential when precision really matters.
| Situation | Appropriate method |
|---|---|
| Non-functional surfaces, Ra 6.3 µm or greater | visual and tactile comparison with the samples |
| In-machine inspection during the production run | comparison, plus a portable roughness tester every now and then |
| Functional surfaces with a Ra of 1.6 µm or less | roughness meter |
| Seats, bearing seats, surfaces marked with “max” or with Rz, Rmax | roughness meter, with a cutoff point and carefully selected points |
| Testing, first workpiece, dispute with the customer | calibrated roughness meter, with stored measurement ratio |
| Very smooth surfaces (Ra below 0.05 µm), soft materials, 3D analysis | noncontact instruments (paragraph 36.13) |
Guideline table prepared by CNC Simulator Italia: Each company establishes its own rules in its control plan (Chapter 38).
36.2 Comparative roughness samples
A comparative sample (known in the workshop as a “roughness plate”) is a plate with a series of surfaces of known roughness, produced by a specific machining process. The most common series reproduce turning, peripheral milling, face milling, and flat and cylindrical grinding; there are also series for lapping, EDM, and sandblasting. The surfaces are usually reproduced by electroplating (with nickel) from a machined and measured original, so all the plates in a series are identical.
How to Use Them
- Choose a series that matches the machining operations of the workpiece: comparing a turned workpiece with a ground sample makes no sense, because the eye and the finger primarily perceive the shape of the grooves;
- Clean the workpiece and place it next to the sample, with the grooves facing the same direction, under the same light—preferably at a glancing angle;
- First, look at it with the naked eye, then with a magnifying glass that magnifies 5 or 10 times;
- Run your fingernail (or fingertip) over the workpiece and the sample, always perpendicular to the grooves and at the same speed: the fingernail vibrates more on rough surfaces;
- Find the two samples that “enclose” the workpiece and write the result as a range: “between Ra 0.8 and Ra 1.6.”
The Limits
The comparison is subjective. With a little practice, you can confidently distinguish a step in the scale (a factor of 2 in Ra); distinguishing finer steps is not realistic. The material can be misleading: a workpiece of polished aluminum feels smoother than a nickel sample with the same Ra value, while cast iron feels rougher. Below an Ra of 0.4 µm, the sense of touch can no longer distinguish well. This comparison does not provide Rz or Rmax values and cannot detect a single deep scratch. Finally, the samples get scratched and dirty: they should be kept in their case and replaced when they show signs of wear. For these reasons, the comparison is an excellent production control, but it is not suitable as a test for a functional surface.
36.3 The stylus roughness tester
The most widely used instrument is the stylus-type roughness tester (also known as a contact or probe-type roughness tester). The principle is simple: an extremely fine tip is dragged across the surface at a constant speed, and a transducer records how much it rises and falls. The signal is amplified, converted into numerical values, filtered, and transformed into the parameters described in Chapter 16.
- Stylus: a small diamond cone with an angle of 60° or 90°, ending in a spherical tip with a nominal radius of 2 µm or 5 µm (10 µm for very coarse surfaces). Diamond does not wear down easily and does not scratch steel, because the measuring force is minimal: less than one millinewton for the 2 µm tip.
- Transducer: converts the vertical displacement of the probe into an electrical signal. The most common type is inductive; laboratory instruments also use optical interference transducers. Vertical resolution reaches the nanometer or a few nanometers, with a range of a few hundred micrometers.
- Shoe or external reference point: provides the line against which toe movements are measured (paragraph 36.4).
- Translation unit: a motor with a precision guide that moves the probe at a constant speed, usually between a few tenths of a millimeter and 1 mm per second. An encoder indicates the probe tip’s location at all times.
- Electronics: amplifies the signal, samples it at a constant rate, applies filters, calculates parameters, displays the profile, and generates the report. In portable devices, it is housed in a unit with a display; in benchtop instruments, it is a PC.
Angles, forces, speeds, and resolutions are approximate values: the manufacturer's specifications and ISO 3274, which describes stylus-type instruments, are authoritative.
The Cape and the Narrow Valleys
The tip has finite dimensions, so it does not fit everywhere: in a valley narrower than its diameter, it stops before reaching the bottom. The roughness tester always measures the profile “seen” by the tip, which is slightly rounded in the valleys. It is a mechanical filter, which is why the standard links the tip radius to the cutoff (table in paragraph 36.6). A chipped or worn tip, which has a radius larger than the nominal value, measures roughness values lower than the actual values.
A groove that is w = 6 µm wide and 3 µm deep. How much does a 5 µm tip sink? And a 2 µm tip?
36.4 Skate or external reference
To measure how much the tip rises and falls, you need a reference line. There are two options, each with different uses.
The Skate
The glide is a rounded support with a radius much larger than that of the tip, mounted next to the tip (in front, behind, or to the side). It glides over the peaks and follows the shape and long waves of the workpiece, so the tip measures only the deviations relative to the glide. The skid-type probe is compact, does not require careful leveling, and is used in portable roughness testers. However, it has some limitations:
- It measures only roughness: it does not provide data on undulation, shape, steps, or the heights of a shoulder;
- If the wave has a length close to the distance between the tip and the tail, the tail rises while the tip drops, and the profile becomes distorted;
- If there is a hole or an isolated peak beneath the skid, a defect appears in the profile that does not actually exist;
- On short curved surfaces (fillets, small spheres), the glider does not sit properly.
The external reference
In benchtop instruments—and in some high-end portable models—the probe is carried by a carriage that slides along a precision linear guide. The instrument records the complete primary profile and then, using filters, calculates roughness, waviness, and form. This is the most accurate method and allows for the measurement of steps, fillets, and profiles as well, but it requires careful leveling of the workpiece (section 36.8) and a stable workbench.
36.5 The Filter: Profile, Waviness, and Roughness
The recorded profile contains everything: the shape of the workpiece, the long waves caused by vibrations and deflections, and the short tool marks (section 16.1). To calculate roughness, only the latter need be considered. The instrument does this using two filters:
- The λs filter removes very short-period irregularities (noise, details finer than the tip can resolve) and yields the primary profile P;
- The λc filter (the cutoff) calculates the average wavy line of the primary profile: this line is the waviness W; what remains, P − W, is the roughness profile R.
The Gaussian Filter
Today, the standard filter is the Gaussian filter (ISO 16610-21). For each point in the profile, it calculates a weighted average of the neighboring points, using bell-shaped weights: nearby points carry a lot of weight, while distant points carry almost none. The width of the bell curve is related to the cutoff. A wave much shorter than λc is “smoothed out” by the average and ends up entirely in the roughness; a wave much longer than λc passes into the mean line and ends up in the undulation. A wave exactly λc long is split in half.
It is measured at λc = 0.8 mm. What fraction of three different waves is accounted for by the roughness?
36.6 Cut-off and Measurement Lengths
Choosing the cutoff is the most important decision before taking a measurement. A cutoff that is too short cuts off part of the roughness and yields values that are too low; one that is too long allows waviness to be included in the roughness and yields values that are too high. The table in Section 16.2 listed cutoff values as a function of Ra; here we supplement it with Rz and the groove width Rsm, which is used for periodic profiles.
| Cut-off λc = lr (mm) | Expected Ra (µm) | Expected Rz (µm) | Rsm, periodic profiles (mm) | ln (mm) |
|---|---|---|---|---|
| 0,08 | more than 0.006 up to 0.02 | more than 0.025 up to 0.1 | more than 0.013 up to 0.04 | 0,4 |
| 0,25 | more than 0.02 up to 0.1 | more than 0.1 up to 0.5 | more than 0.04 up to 0.13 | 1,25 |
| 0,8 | more than 0.1 up to 2 | more than 0.5 up to 10 | more than 0.13 up to 0.4 | 4 |
| 2,5 | more than 2 and up to 10 | more than 10 up to 50 | more than 0.4 up to 1.3 | 12,5 |
| 8 | more than 10 up to 80 | more than 50 up to 200 | more than 1.3 up to 4 | 40 |
Guideline ranges taken from ISO 4288 (now incorporated into ISO 21920-3), to be used when the drawing does not specify a cutoff value. If the drawing specifies a cutoff value, the drawing takes precedence.
How to Choose
- Periodic profile (turning, milling with regular pitch): Estimate Rsm, which for turning is nearly equal to the feed rate f, and select the cutoff value from the Rsm column. The bands are designed so that each base length contains at least two grooves.
- Non-periodic profiling (grinding, lapping, EDM, sandblasting): Estimate Ra (from the drawing or by comparison with the reference plate), select the cut-off value, measure, and verify that the measured value falls within the range defined by that cut-off. If it falls outside the range, change the cutoff value and repeat the process.
The tip and sampling pitch must also be compatible with the cutoff frequency. ISO 3274 specifies the following combinations:
| λc (mm) | λs (µm) | tip radius (µm) | Maximum sampling step Δx (µm) |
|---|---|---|---|
| 0,08 | 2,5 | 2 | 0,5 |
| 0,25 | 2,5 | 2 | 0,5 |
| 0,8 | 2,5 | 2 | 0,5 |
| 2,5 | 8 | 5 | 1,5 |
| 8 | 25 | 10 | 5 |
Approximate values according to ISO 3274. In practice, with λc = 0.8 mm, the 5 µm stylus is also often used, with a negligible error when Ra exceeds approximately 0.5 µm.
Base Length, Measurement Length, and Total Travel Length
The base length lr is equal to the cutoff; the evaluation length ln, unless otherwise specified in the drawing, is five base lengths (Section 16.2). However, the probe travels a longer distance: the Gaussian filter, in order to calculate the average line at the beginning and end, requires data points both before and after those points. Therefore, it travels a start-up stroke and an exit stroke, which are not included in the calculation.
A pin is turned at f = 0.2 mm/rev and rε = 0.8 mm; the drawing specifies Ra 1.6 µm. What cutoff value should be used?
A bore honed to a Ra of 0.2 µm and a hub rough-turned on a lathe with a feed rate of f = 0.5 mm/rev and a radius of 1.6 mm.
With λc = 0.8 mm, a Gaussian filter, and a speed of 0.5 mm/s. And what about the old 2RC filter?
A turned shoulder with a Ra of 1.6 µm is only 3 mm wide. Can it be measured with λc = 0.8 mm?
36.7 Direction and Position of the Measure
Perpendicular to the furrows
Unless otherwise specified in the drawing, the measurement is taken perpendicular to the direction of the grooves (the direction of the grooves on the symbol is explained in paragraph 16.6). In this direction, the probe passes through all the grooves, and the measured roughness is the highest: this is the worst-case scenario, the one that matters for the function. When measuring along a groove, the tip slides along the bottom and records only the small irregularities of the cutting edge: the number is much lower and makes a surface appear good when it is not.
On a turned workpiece, the grooves run around the axis, so the correct travel direction is along the generatrix, parallel to the axis. On a faced surface, the grooves are circles (or a spiral), and the correct travel direction is radial. On a face-milled surface, the grooves are arcs: the measurement is taken in the direction of feed, at the center of the pass. Surfaces without a defined direction (sandblasting, erosion, cross-lapping) can be measured in any direction; as a precaution, measurements should be taken in two directions at 90°.
Where and how often
- Measure where you expect the worst value: experience and the reference plate will help you find it (usually near the shoulders, where the tool slows down, or where the workpiece is more flexible and vibrates);
- On a uniformly finished surface, start with one measurement and increase it in accordance with the acceptance rule (paragraph 36.10);
- On shafts, measurements are taken along multiple generatrices—for example, three at 120° intervals—at the beginning, middle, and end of the section;
- For small surfaces, the number of base lengths is reduced, and this is stated (example 36.7);
- In holes, probes with long, thin arms are used; with special tips, they can enter holes just a few millimeters in diameter, but the depth they can reach is limited;
- For radii and curved surfaces, an external reference and a filter that removes the shape are used; with the probe, measurements are taken only along a straight generatrix;
- Keep at least a few tenths of a millimeter away from edges, chamfers, holes, and edges: if the probe touches an edge, it will give nonsensical readings.
36.8 Step-by-Step Implementation
A roughness measurement takes only a few seconds, but preparation is more important than the measurement itself. Here is the procedure to follow, whether using a portable or benchtop roughness tester.
- Cleaning. Remove oil, emulsion, chips, and dust using a lint-free cloth and a suitable solvent. Never clean with sandpaper or a wire brush: this alters the surface you intend to measure. A speck of dust under the tip becomes a peak of a few micrometers.
- Instrument Check. At the beginning of the shift (or before an important series of measurements), measure the supplied reference sample and compare the value with that on its certificate (paragraph 36.14). If the deviation exceeds the specified limit, adjust the instrument or send it for calibration.
- Support. The workpiece must be held steady: on a flat surface, in a vise, or on a V-block for shafts. Keep it away from presses, compressors, and rough-machining tools: floor vibrations will be transmitted to the workpiece. The portable machine rests on its own feet or on a V-block; do not hold it in your hand while it is traveling.
- Leveling and Positioning of the Probe. The travel must be parallel to the surface so that the probe tip remains within its measuring range. Using the external reference, level the workpiece (or table) until the position indicator remains centered throughout the entire travel; when using the slide, simply ensure that the probe is positioned straight and that the probe tip starts at the midpoint of the range.
- Selection of parameters. Cut-off and number of base lengths (section 36.6), Gaussian filter, parameters to be calculated (those from the drawing, plus Rz and Rt, which cost nothing), acceptance rule.
- Measurement. Start the travel without touching either the workpiece or the workbench.
- Read and check the profile. Before entering the number, look at the profile: a very high isolated peak indicates dust or a burr; a profile that rises entirely on one side indicates improper leveling; a long, regular wave may indicate vibration. If the profile looks suspicious, clean it and repeat the process.
- Repeat and record. Take measurements at the other designated points, note the values along with their locations, and save the report.
36.9 The measurement ratio and how to interpret it
The result of a measurement is a report (displayed on-screen, printed, or saved as a file) consisting of three parts: the measurement conditions, the profile graph, and the parameter values. Without the conditions, the numbers cannot be compared to anything.
- Conditions: cut-off values for λc and λs, number of base lengths and evaluation length, filter, nose radius and angle, slide or external reference, speed, date, tool, operator, position on the workpiece.
- Profile: Keep in mind that the two scales are different. In the figure, the vertical magnification is about 90 times that of the horizontal magnification (compare the two scale bars, 2 µm and 0.5 mm). The grooves look like sharp saw teeth, but in reality they are very wide and very shallow depressions.
- Ra, Rq, Rz, Rt: defined in Chapter 16 (Sections 16.3 and 16.4).
- Rmax (or Rz1max): the highest of the five Zt heights, useful for sealing surfaces.
- Rsm: the average width of the profile elements, that is, the average distance between one peak and the next, in mm. On a turned workpiece, it is almost equal to the feed rate: in the figure, 0.150 mm with f = 0.15 mm/rev. If Rsm differs from the feed rate, something is wrong (vibrations, cutting edge).
- Rmr(c): the lift-to-drag ratio, explained below.
How the tool calculates the parameters
To understand how the electronics work, let’s redo the calculation by hand for a few points. The instrument uses thousands of them, but the procedure is the same: find the average line, measure the ordinates from that line, and apply the definitions from Chapter 16.
The transducer provides sixteen readings at constant intervals, measured from any zero point (in µm): 1.9 · 3.2 · 0.7 · −0.8 · 1.1 · 2.5 · −0.3 · −2.0 · 0.4 · 2.2 · 3.4 · 0.8 · −1.1 · 0.1 · 1.5 · −0.8. Calculate Ra, Rp, Rv, and Rt for this section.
Across the five base lengths, the roughness meter detects these highest peaks (Zp) and deepest valleys (Zv) (in µm, from the mean line):
| Section | 1 | 2 | 3 | 4 | 5 |
|---|---|---|---|---|---|
| Zp | 2,6 | 3,1 | 2,4 | 3,5 | 2,8 |
| Zv | 2,9 | 2,5 | 3,3 | 2,7 | 3,6 |
The ratio shown in Figure 36.10 is based on a turned shaft with f = 0.15 mm/rev and rε = 0.8 mm. Compare the measured values with the theoretical values from the formula in Section 26.10.
The Abbott Curve and the Lift-to-Drag Ratio
Let’s imagine cutting through the profile with a horizontal line at a depth c below the highest peak. The line passes through sections of material and sections of void. The sum of the sections of material, divided by the evaluation length, is the lift ratio Rmr(c): It indicates how much surface area a resting plane would touch after the peaks have been worn down by c. By varying c from the highest peak to the deepest valley, we obtain the Abbott curve (or lift curve), which ranges from 0% to 100%.
Using the ordinates from Example 36.8 (highest peak 2.6 µm), calculate Rmr for c = 1 µm and c = 3 µm.
Two surfaces with the same Ra value can have very different Abbott curves (Figure 16.6): a surface with plateaus and narrow valleys has a high bearing ratio even just a few micrometers from the peak, carries the load well, and retains oil in the valleys. For these surfaces, such as cylinder liners and guides, the standard also defines the parameters Rpk (height of the peaks that will wear down during break-in), Rk (depth of the load-bearing core), and Rvk (depth of the valleys that retain the lubricant), derived from the Abbott curve. Here, it is sufficient to know that these parameters exist and that they can be found in the report when specified in the drawing.
36.10 Comparison with the drawing: the 16% rule and the maximum rule
A surface does not have “a single” roughness value: measurements taken at different points yield different values. To determine whether the workpiece is in compliance, a rule is needed, which depends on how the value is specified in the drawing (section 16.6).
- 16% Rule (the default rule in ISO 1302 and ISO 4288, when only “Ra 1.6” is specified): The surface is acceptable if no more than 16% of the measured values exceed the limit.
- Maximum Rule (when it says “Ra max 1.6”): No measured value, across the entire surface, must exceed the limit. Measurements must be taken at least three times at the points where the highest values are expected.
With the ISO 21920 series, the acceptance criteria have been revised: for new drawings, it is advisable to check the standard referenced in the title block and, if in doubt, ask the engineering department which criterion applies. Here we describe the two standard criteria, which are the ones found on the vast majority of drawings.
The 16% Rule Procedure
ISO 4288 provides a step-by-step procedure that omits certain measurements when the surface is clearly in good condition. In simplified form:
- Take an initial measurement at the point that appears to be the worst: if the value does not exceed 70% of the limit, the surface is accepted;
- Otherwise, two more measurements are taken: if none of the three values exceeds the limit, the surface is accepted; if two or more values exceed the limit, it is rejected;
- If even one value exceeds the limit, the total number of measurements reaches twelve: the area is accepted if no more than two of the twelve values exceed the limit.
Educational summary prepared by CNC Simulator Italia; for contractual acceptance testing, the text of the standard referenced on the drawing shall prevail.
A shaft has a designed Ra of 0.8 µm, with the 16% rule. 70% of the limit is 0.7 × 0.8 = 0.56 µm. Three different workpieces.
On a surface with a surface roughness (Ra) of 1.6 µm, twelve values (in µm) are measured: 1.32 · 1.45 · 1.18 · 1.52 · 1.71 · 1.39 · 1.27 · 1.48 · 1.66 · 1.35 · 1.41 · 1.29. Does the surface meet the specifications?
The same twelve measurements, but the drawing says “Ra max 1.6.” Then there is another workpiece marked “Ra max 0.8,” measured with an expanded uncertainty U = 0.06 µm, which gives a maximum of 0.71 µm; a third workpiece gives a maximum of 0.77 µm.
36.11 Common Errors
Almost all errors in roughness measurements have the same effect: a number that appears precise—with three digits—but does not accurately describe the surface. These are the most common ones.
| Error | What's happening? | Remedy |
|---|---|---|
| Measurement parallel to the furrows | values that are much lower than the actual values (Figure 36.8) | perpendicular to the grooves, unless otherwise specified in the drawing |
| Worn or chipped tip | The actual radius increases, the tip does not enter the valleys, and Ra and Rz decrease; chipping may also cause scratches on the workpiece | Periodic inspection of the tip using the sample or under a microscope; replacement |
| Vibrations | Regular waves or noise in the profile, high values | Damped bench, installed away from heavy machinery, no contact during travels |
| Skating on a curved or short surface | The skate rises and falls with the curve; the profile deforms. | external reference with a shape filter, or travels along a straight generatrix |
| Cut-off is too short | Part of the roughness is accounted for by waviness: low values | Selecting the cutoff value using the table in paragraph 36.6 and verifying the range |
| Cut-off is too long | Waviness contributes to roughness: high values | as above |
| Undeclared reduced evaluation length | less stable values that cannot be compared with other measurements | Always include the number of base lengths on the report |
| Dirty or oily workpiece | Isolated peaks (powder) or filled valleys (oil): distorted Rz and Rt values | Thorough cleaning; check the profile before reading the numbers |
| Travel that touches an edge or a hole | huge spikes in the graph, nonsensical values | to start and end far from the edges |
| Off-screen tip due to incorrect leveling | profile cut at the top or bottom, error message, or low values | Level the workpiece; center the drill bit within the area |
| A filter different from the reference filter (2RC instead of Gaussian) | differences of a few percent—sometimes more—compared to the customer's measurements | Use the Gaussian filter and include it in the report |
The hub from Example 36.4 (theoretical Ra of 5.0 µm, grooves every 0.5 mm) is routinely measured with λc = 0.8 mm instead of 2.5 mm. What happens to the main component of the profile—the one corresponding to the groove pitch?
36.12 Portable, benchtop, and in-process roughness meters
Laptops
The portable roughness tester features a translation unit and a sliding probe in a small, battery-powered housing that can often be detached from the display. It is taken to the machine, placed on the workpiece—even while it is still mounted—measures Ra, Rz, Rt, Rmax, Rsm, and a few other parameters, and prints or saves the results. It is the workshop’s go-to tool: fast, rugged, and with a measurement uncertainty suitable for most inspections. With magnetic mounts, V-blocks, and extensions, it adapts to holes, grooves, and vertical surfaces.
Countertop
The benchtop roughness tester (or profilometer) features a motorized column, a table that can often be tilted for leveling, long travels (up to tens or hundreds of millimeters), and an external reference. It measures the complete primary profile—including waviness, shape, radii, and angles—and calculates all parameters, including the Abbott curve and Rk parameters. It is housed in the metrology lab on a damped bench and serves as the instrument for testing and dispute resolution.
In the car and in production
- Roughness probes on coordinate measuring machines (Chapter 37): The machine positions the probe at the correct location and orients it perpendicular to the grooves, even inside holes; dimensions and roughness are measured with a single setup.
- Roughness probes in machine tools: Mounted in the spindle like a measuring probe, they inspect the freshly machined surface before the workpiece is removed. They require a stationary, clean machine and a dedicated measurement cycle.
- Optical light-scattering sensors: A beam strikes the surface, and a sensor measures how the reflected light scatters. In a fraction of a second, they provide a roughness index without making contact; they are used to inspect 100% of workpieces on the production line, after comparing the index with measurements taken using a stylus.
- Statistical control: The values measured during production are recorded on control charts (Chapter 38). A slowly increasing Ra value indicates tool wear; a sudden jump indicates chipping or a change in material.
36.13 3D Measurements and Noncontact Methods
From the profile to the surface
A profile is a line: it describes the surface only along the path of the probe tip. A ground, sandblasted, or eroded surface may have isolated defects (a crater, a scratch) that one travel detects but another does not. Area measurements, on the other hand, cover an entire area—for example, a square a few millimeters in size—much like a map of heights. The parameters are defined by ISO 25178 and are identified by the letter S:
- Sa: arithmetic mean of the absolute heights over the area, the areal “counterpart” of Ra;
- Sq: standard deviation of the heights (same as Rq);
- Sz: maximum height of the surface, from the highest peak to the deepest valley in the entire area;
- Other parameters describe the direction of the grooves, the actual surface area, and the volumes of material and void space, which are important for sealing and lubrication.
Sa and Ra are not the same thing: on surfaces without a specific direction, they are similar; on surfaces with oriented grooves, Ra depends on the direction of travel, while Sa does not. Different parameters are specified on the drawing, and they cannot be interchanged without the designer’s approval.
A 4 × 4 grid of heights (µm), measured from any plane: first row 0.6 · −0.4 · 1.2 · 0.2; second row −0.9 · 0.3 · −0.2 · 1.0; third row 0.5 · −1.3 · 0.8 · −0.5; fourth row 0.1 · 0.9 · −0.7 · 0.4.
Non-contact methods
For area measurements, for very fine or soft surfaces (plastics, coatings, lenses), and for microstructures, non-contact optical instruments are used. The three most common principles are as follows.
- Confocal microscope: A small hole (pinhole) in front of the sensor allows only light from the points in focus to pass through. By moving the objective vertically, the position of maximum brightness is determined for each point, which corresponds to its height. In the color version, white light is used along with a lens that focuses each color at a different dimension: the color that is reflected back indicates the height, without moving anything.
- White-light interferometry: the light splits into two beams, one of which strikes a workpiece and the other a reference mirror; when they recombine, they form light and dark fringes. With white light, the fringes are sharp only when the two paths are equal: during vertical scanning, the point of maximum contrast provides the dimensions of each pixel. The vertical resolution is less than one nanometer.
- Focus-variation microscope: It takes many images at different dimensions and, for each pixel, searches for the image in which it is sharpest. It works well on rough and sloped surfaces, but it needs some roughness; on a mirror-smooth surface, it can’t find anything to focus on.
| Method | Strengths | Limitations |
|---|---|---|
| Stilo | complies with standards and is comparable, sturdy, affordable, works well in the workshop | only profiles (or slow-scanning areas); a tip that filters and can scratch soft materials |
| Confocal | Glossy and matte surfaces, steep slopes, high lateral resolution | small field of view, cost, susceptible to reflections |
| White-Light Interferometry | Extremely high vertical resolution, fast on fine surfaces | Difficult to use on very rough or very steep surfaces; sensitive to vibrations |
| Focus Shift | rough surfaces, shapes and edges, tools | Not suitable for mirror-like surfaces; lower vertical resolution |
Indicative comparison prepared by CNC Simulator Italia. Optical instruments and styluses do not always yield the same reading on the same surface: the probe and the optical system “see” differently. For inspection purposes, use the method specified in the drawing or agreed upon with the customer. Other modern measuring instruments are covered in Chapter 37.
36.14 Calibration of the roughness tester
Like any measuring instrument (paragraph 31.14), the roughness tester must be calibrated at regular intervals and checked frequently. Roughness calibration standards, as described in ISO 5436-1, are used and classified into types based on what they are designed to measure.
- Type A: Wide grooves with a flat (or rounded) bottom and a certified depth. They are used to calibrate the vertical amplification: the instrument must read the depth d.
- Type B: Narrow grooves or thin edges, to check the condition and radius of the tip.
- Type C: periodic waveforms (sine waves, triangles, arcs) with known period and amplitude; they also control the horizontal scale, the filter, and the calculation of Ra and Rsm.
- Type D: irregular profiles, similar to those produced by grinding, that repeat every base length; they certify Ra and Rz and verify the instrument under conditions that most closely resemble actual use.
- Type E: Shape gauges (arcs, prisms) for instruments that also measure profiles and radii.
True calibration is performed in an accredited laboratory, which issues a certificate listing the measured values and their uncertainty. In the workshop, every time the instrument is turned on or at the start of a shift, a check is performed using the calibration block supplied with the instrument (usually a block with a Ra of a few micrometers and its certified value): if the reading deviates too much, adjust the instrument’s gain according to its manual, or send it in for calibration. The standard must be kept clean, in its case, and always measured at the same point and in the same direction: the tip, due to repeated use, will eventually wear it down.
The supplied sample has a certified Ra value of 2.97 µm. The company allows for a tolerance of ±5% (internal standard). Five readings taken this morning: 3.04 · 3.01 · 3.07 · 2.99 · 3.05 µm.
36.15 The Formulas in This Chapter
| What | Formula or rule | Units |
|---|---|---|
| Ra scores in league play | Ra = (|z1| + ... + |zn|) / n, where z is the mean | µm |
| Rz, Rmax, Rt (Chapter 16) | Rz = average of the 5 Zt values · Rmax = max(Zt) · Rt = Zp,max + Zv,max | µm |
| Sinking of the tip | h = r − √(r² − (w/2)²) | µm |
| Gaussian Filter Transfer Function | TW = 0.5(λc/λ)² · TR = 1 − TW | pure number |
| Evaluation Length | ln = 5 · lr = 5 · λc | mm |
| Total Travel (Gaussian Filter) | lt = ln + λc = 6 · λc | mm |
| Number of points and time | N = ln / Δx · t = lt / v | points · s |
| Lift-to-drag ratio | Rmr(c) = Ml(c) / ln × 100 | % |
| Theoretical Roughness in Turning (Chapter 26) | Ra ≈ 1000 · f² / (31.2 · rε) · Rt = 1000 · f² / (8 · rε) | µm |
| The 16% Rule (statistical interpretation) | no more than 2 of the 12 values above the limit · approximately mean + s ≤ limit | µm |
| Compliance with Uncertainty | value + U ≤ upper limit | µm |
| Area parameter Sa | Sa = (|z1| + ... + |zn|) / n, where z is from the mean plane | µm |
| Validation Error | e = (average − certified value) / certified value × 100 | % |
36.16 Exercises
Try to solve them on your own, using a calculator, before looking at the solution.
Solution
Sum 3.6 → average line 3.6 / 12 = 0.3 µm. Y-coordinates: 0.6 · −0.7 · 1.4 · 0 · −1.5 · 0.3 · 1.8 · −1.2 · −0.1 · 1.1 · −1.8 · 0.1. Sum of absolute values 10.6 → Ra = 10.6 / 12 = 0.88 µm. Rt = 1.8 + 1.8 = 3.6 µm.
Solution
Zt = 3.9 × 3.9 × 4.0 × 4.4 × 4.6 µm. Rz = 20.8 / 5 = 4.16 µm. Rmax = 4.6 µm. Rt = 2.5 + 2.6 = 5.1 µm.
Solution
Lapping: range “greater than 0.02 up to 0.1” → 0.25 mm. Grinding (Ra 0.1): still within the range “greater than 0.02 up to 0.1” → 0.25 mm (pay attention to the edge). Turning: Theoretical Ra = 0.3² · 1000 / (31.2 · 1.6) = 90 / 49.92 = 1.80 µm; Rsm ≈ 0.3 mm in the range “over 0.13 up to 0.4” → 0.8 mm, confirmed by Ra 1.8 µm in the range “over 0.1 up to 2.” Non-periodic milling Ra 3.2: 2.5 mm.
Solution
ln = 5 × 2.5 = 12.5 mm; lt = 6 × 2.5 = 15 mm; t = 15 / 1 = 15 s. With the 2RC: lt = 7 × 2.5 = 17.5 mm, 17.5 s.
Solution
ln = 5 × 0.25 = 1.25 mm = 1250 µm; N = 1250 / 0.5 = 2500 points.
Solution
1.6 mm wavelength: λc / λ = 0.5 → TW = 0.5 × 0.25 = 0.841 → TR = 15.9%. 0.4 mm wavelength: λc / λ = 2 → TW = 0.54 = 0.0625 → TR = 93.75%.
Solution
w / 2 = 1.5 µm. 2 µm tip: h = 2 − √(4 − 2.25) = 2 − 1.323 = 0.68 µm. 5 µm tip: h = 5 − √(25 − 2.25) = 5 − 4.770 = 0.23 µm. Both remain suspended: no tip can measure such a narrow groove in its entirety.
Solution
Above 3.2, there are 3.34, 3.41, and 3.27: three values out of twelve, more than the two allowed → non-conforming. Mean = 2.999 µm, s = 0.250 µm, mean + s = 3.249 µm > 3.2: statistical analysis also indicates that the 16% limit has been exceeded.
Solution
All values are below 0.4, but the highest one—0.38 + 0.03 = 0.41 µm—exceeds the limit: compliance is not proven, and the value falls within the zone of doubt. Compliance is proven if all values are below 0.4 − 0.03 = 0.37 µm.
Solution
Theoretical Ra = 0.12² · 1000 / (31.2 · 0.4) = 14.4 / 12.48 = 1.15 µm; ratio 1.45 / 1.15 = 1.26. Required theoretical Ra = 0.8 / 1.26 = 0.635 µm; f = √(31.2 × 0.4 × 0.635 / 1000) = √0.00792 = 0.089 mm/revolution, or approximately 0.08 mm/revolution.
Solution
Average = 2.11 / 5 = 0.422 µm; error = 0.422 − 0.45 = −0.028 µm = −6.2%. Outside the ±5% range: the instrument is reading too low. Check the tip (a worn tip produces low readings), clean the sample, and if the error persists, adjust the instrument or send it in for calibration.
Solution
Highest peak: 1.6 µm. c = 1 µm → level 0.6 µm: above this are 1.6, 1.2, and 0.8 → 3 out of 10 → RMR = 30%. c = 2 µm → level −0.4 µm: above this are 1.6, 1.2, 0.8, 0.5, 0.2, and −0.3 → 6 out of 10 → Rmr = 60%.
Solution
Space for the bases: 2 − 0.8 = 1.2 mm → 1.2 / 0.8 = 1.5 → a single base length (ln = 0.8 mm, lt = 1.6 mm). With λc = 0.25 mm, the 0.2 mm grooves would be outside the Rsm range for that cutoff (up to 0.13 mm): each base would contain only 0.25 / 0.2 = 1.25 grooves, and the filter would cut out part of the roughness: the value would be too low. Measure using one base, record the result in the report, and repeat the measurement at multiple points along the shoulder.